PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080120 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Published December 24, 2014 · Recall initiated: November 17, 2014
Summary
FDA records medical device Class II recall Z-0593-2015, published December 24, 2014 and initiated November 17, 2014. Recalling firm: Stryker Endoscopy; Status: Terminated; Reported quantity: 180,573 units total; Lot: Serial/Lot 06CON06110 06CON06111 0816588/0816395 0816693/0816395 0913442/0817444 0913984/1014813 0914894/1013609 0914894/1014814 0914894/1016062 0914894/1016825 0915431/1016825 1015779/0914894 1016736/1114541 1016739/1016398 1016739/1113020 1016739/1114541 1016846/1114541 1113331/1115651 1113332/1115651 1114831/1115651 1115971/1213123 1116462/1213123 1116463/1213978 1116463/1213979 1314474/1314476. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Endoscopy
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lot
- Serial/Lot 06CON06110 06CON06111 0816588/0816395 0816693/0816395 0913442/0817444 0913984/1014813 0914894/1013609 0914894/1014814 0914894/1016062 0914894/1016825 0915431/1016825 1015779/0914894 1016736/1114541 1016739/1016398 1016739/1113020 1016739/1114541 1016846/1114541 1113331/1115651 1113332/1115651 1114831/1115651 1115971/1213123 1116462/1213123 1116463/1213978 1116463/1213979 1314474/1314476
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,December 24, 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, BIPOLAR FORCEPS, MICRO TIP, 45CM, P/N 0250080115 nationwide in US and worldwide: Algeria, Argentina,December 24, 2014
- FDAPKG, BIPOLAR FORCEPS, PADDLE, 33CM, P/N 0250080121 Laparoscopic Manual Instruments are intended for cutting,December 24, 2014
Questions about this recall
Was PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080120 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. recalled?
Yes. FDA published recall Z-0593-2015 on December 24, 2014. Reported hazard: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Voluntary: Firm initiated · Letter