Catalog No. 9251R1, 9251R2, 9251R3 TRACHEOSTOMY CARE SET / STERILE - LATEX FREE
Published January 10, 2024 · Recall initiated: November 30, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm is recalling all kits containing recalled lots of Nurse Assist Sterile 0.9% Normal Saline. Nurse Assist identified product in distribution for which the required sterility assurance level (SAL) of 10^-6 cannot be guaranteed. Risks to patients if product is used include blood stream infections, urinary tract infections, or respiratory infections; in the worst-case scenario, there is risk of sepsis or death.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Busse Hospital Disposables, Inc.
- Distributor
- US Nationwide distribution.
- Lot
- UDI-DI 00849233011069 Lots 2210087 2210165 2310361
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Sell products like this?