Back to search
FDAFDA · Food and drugs·Recall number Z-0608-2021

Flexor Check-Flo Introducer, RPN KCFW-8.0-35-70-RB-HFANL0-HC, Order Number G56223 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.

Published December 30, 2020 · Recall initiated: November 24, 2020

Summary

FDA records medical device Class I recall Z-0608-2021, published December 30, 2020 and initiated November 24, 2020. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 228; Lot: Lot Numbers: 13297248, 13318063, 13320595, 13334764, 13388304, 13435140, 13291880X, NS13246556, NS13265340, NS13288123. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cook Inc.
Distributor
US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS, MI, ND, NY, SC, VT, CO, HI, KY, MN, NE, OH, SD, WA and Global Distribution to: AU, SG, Austria, Belgium, Canary Islands, CN, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Italy, KR, Latvia, Lithuania, MY, Morocco, Netherlands, Norway, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, TW, TH, and United Arabs Emirates.
Lot
Lot Numbers: 13297248, 13318063, 13320595, 13334764, 13388304, 13435140, 13291880X, NS13246556, NS13265340, NS13288123
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Flexor Check-Flo Introducer, RPN KCFW-8.0-35-70-RB-HFANL0-HC, Order Number G56223 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature. recalled?

Yes. FDA published recall Z-0608-2021 on December 30, 2020. Reported hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.

What should I do if I have this product?

Voluntary: Firm initiated · Letter