Model 208-70-110, Long Thread Tiger Large Cannulated Headless Screw, 7 x 110 mm labeled as "7.0 x 110mm LTTHCS REF 208-70-110 QTY 30 LOT TSL003721 2016-03-29 Trilliant Surgical, LTD"
Published November 23, 2016 · Recall initiated: September 2, 2016
Summary
FDA records medical device Class II recall Z-0609-2017, published November 23, 2016 and initiated September 2, 2016. Recalling firm: Trilliant Surgical Ltd.; Status: Terminated; Reported quantity: 11; Lot: Lot TSL003721. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Screws were identified as having the incorrect part number and lot number marked upon them. The part number incorrectly indicated a length of 115 mm rather than 110 mm. The lot number incorrectly indicated lot number of TSL003765 rather than TSL003721.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Trilliant Surgical Ltd.
- Distributor
- CA, AZ, WI, FL, NJ, OH
- Lot
- Lot TSL003721
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATiger Screw Sterilization Tray Base Assembly Model 210-00-001 Revision N, O The product is a component of aJanuary 18, 2017
- FDASOMATOM Definition Flash, Computed tomography x-ray system The Siemens SOMATOM Definition AS/ AS+ (ProjectNovember 23, 2016
- FDAOMNIBotics Tracker Kit -Model: RM-10000 The TOTAL KNEE SURGETICS Navigation System with iBlock (NowNovember 23, 2016
- FDASOMATOM Definition Edge, Computed tomography x-ray system The Siemens SOMATOM Definition AS/ AS+ (Project P46)November 23, 2016
- FDAiConnect Access software. The firm name on the label is Merge Healthcare, Hartland, WI.November 23, 2016
- FDASOMATOM Definition Flash, Computed tomography system Product Usage: SOMATOM Definition AS systems are intendedNovember 23, 2016
Questions about this recall
Was Model 208-70-110, Long Thread Tiger Large Cannulated Headless Screw, 7 x 110 mm labeled as "7.0 x 110mm LTTHCS REF 208-70-110 QTY 30 LOT TSL003721 2016-03-29 Trilliant Surgical, LTD" recalled?
Yes. FDA published recall Z-0609-2017 on November 23, 2016. Reported hazard: Screws were identified as having the incorrect part number and lot number marked upon them. The part number incorrectly indicated a length of 115 mm rather than 110 mm. The lot number incorrectly indicated lot number of TSL003765 rather than TSL003721.
What should I do if I have this product?
Voluntary: Firm initiated · Letter