C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Published December 18, 2019 · Recall initiated: October 21, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Integra LifeSciences Corp.
Distributor
US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK
Lot
OUS Ops Manual P/N 60905217
Risk classification
Class II