S3 Med/Surg Bed Secure II and the 3002 S3 are the beds typically found in the Med-Surg units of the hospitals. The bed is designed for a 15 year expected service life under normal use conditions, and with appropriate periodic maintenance as described in the maintenance manual for each device. There are minimal differences between the functionality of the two models, each of which function as a highly complex system of a large number of individual components.
Published January 8, 2014 · Recall initiated: November 26, 2013
Summary
FDA records medical device Class II recall Z-0618-2014, published January 8, 2014 and initiated November 26, 2013. Recalling firm: Stryker Medical Division of Stryker Corporation; Status: Terminated; Reported quantity: 108,769 Total; Lot: Models: 3002 S3 EX, 3002 S3 PATRIOT, 3002 S3 CUSTOM, 3002 S3 ZOOM, and 3002 S3 SIGNATURE. And Serial Numbers:080615942, 080715062, 080916706, 080916707, 080916708, 080916709, 080916710, 080916711, 080916712, 080916713, 080916714, 080916715, 080916716, 080916717, 090415690, 100215206, 100215207, 100215208, 100215209, 100215210, 100215211, 100215212, 100215213, 100215214, 100215215, 100215216, 100215217, 100215218, 100215219, 100215220, 100215221, 100215222, 100215223, 100215224, 100215225, 100215804, 100215805, 100215806, 100215807, 100215808, 100215809, 100215810, 100315078, 100315079, 100315080, 100315081, 100315082, 100315083, 100315084, 100315085, 100315086, 100315087, 100315088, 100315089, 100315090, 100315091, 100315092, 100315670, 100315671, 100315672, 100315673, 100315674, 100315675, 100315676, 100315677, 100315678, 100315679, 080815116, 091215376, 091215377, 091215378, 110215808, 110215809, 100815273, 100815274, 100815275, 100815276, 100815277, 100815278, 100815279, 100815280, 100815281, 100815282, 100815283, 100815284, 100815285, 100815286, 100815287, 100815288, 100815289, 100815290, 100815291, 100815292, 100815293, 100815294, 100815295, 101116017, 110215554, 100815726, 100815727, 100815728, 100815729, 100815730, 100815731, 100815732, 100815733, 100815734, 100815735, 100815736, 100815737, 100815738, 100815739, 100815740, 100815741, 100815742, 100815743, 100815744, 100815745, 100815746, 100815747, 100815748, 100815749, 100815750, 100815751, 100815752, 100815753, 100815754, 100815755, 100815756, 100815757, 100815758, 100815759, 100815760, 100815761, 100815884, 100815885, 100815886, 100815887, 100815888, 100815889, 100815890, 100815891, 100815892, 100815893, 100815894, 100815895, 100815896, 100815897, 100815898, 100815899, 100815900, 100815901, 100815902, 100815903, 100815904, 100815905, 100815906, 100815907, 100815908, 100815909, 100815910, 100815911, 100815912, 100815913, 100815914, 100815915, 100815916, 100815917, 100815918, 080815227, 090616274, 090616275, 090616276, 090616277, 090616278, 090616279, 080715001, 080715002, 080715003, 080715004, 080715005, 080715006, 080715007, 080715008, 080715009, 080715010, 080715011, 080715012, 080715013, 080715014, 080715015, 080715016, 080715017, 080715018, 080715019, 080715020, 080715021, 080715022, 080715023, 080715024, 080715025, 080715026, 080715027, 080715028, 080715029, 080715030, 080715031, 080715081, 080715082, 080715083, 080715084, 080715085, 080715086, 080715087, 080715088, 080715089, 080715090, 080715091, 080715092, 080715093, 080715094, 080715095, 080715096, 080715097, 080715098, 080715099, 080715100, 080715101, 080715102, 080715103, 080715104, 080715105, 080715106, 080715107, 080715108, 080715109, 080715110, 080715111, 080715114, 081215107, 090316108, 090415132, 090815025, 090815026, 090815027, 090815028, 090815029, 090815030, 090815031, 090815032, 090815033, 090815034, 090815035, 090815036, 090815037, 090815038, 090815039, 090815040, 090815041, 090815042, 090815043, 091116010, 091116011, 091116012, 091116013, 091116014, 091116015, 091116016, 091116017, 091116018, 091116019, 091116020, 091116021, 091116022, 091116023, 091116024, 091116025, 091116026, 091116027, 091116028, 091116029, 091116030, 091116031, 091116032, 091116033, 091116034, 091116035, 091116036, 091116037, 091116038, 091116039, 091116040, 100315027, 100315352, 100416123, 100515548, 100615297, 100615298, 100615300, 100615592, 100615696, 100615699, 100615700, 100615718, 100615829, 100616115, 100815876, 100815878, 100816127, 101015652, 101015653, 101015675, 101015677, 101015679, 101015955, 101015957, 101015958, 101015961, 101015962, 101016094, 101016097, 101016099, 101016100, 101016103, 10. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Stryker Medical has identified a potential issue associated with Secure II and S3 Med/Surg beds built between April 2nd, 2005 and July 26th, 2011. The affected beds have the potential to experience damage to the Foot End Cover and/or Power Coil Cable due to reduced clearance between the Foot End Cover and Foot End Lift Header.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Medical Division of Stryker Corporation
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, Chile, China, Argentina, Brazil, Germany, Korea, Mexico, Singapore, UK, Europe, Middle East, Africa, and Latin America.
- Lot
- Models: 3002 S3 EX, 3002 S3 PATRIOT, 3002 S3 CUSTOM, 3002 S3 ZOOM, and 3002 S3 SIGNATURE. And Serial Numbers:080615942, 080715062, 080916706, 080916707, 080916708, 080916709, 080916710, 080916711, 080916712, 080916713, 080916714, 080916715, 080916716, 080916717, 090415690, 100215206, 100215207, 100215208, 100215209, 100215210, 100215211, 100215212, 100215213, 100215214, 100215215, 100215216, 100215217, 100215218, 100215219, 100215220, 100215221, 100215222, 100215223, 100215224, 100215225, 100215804, 100215805, 100215806, 100215807, 100215808, 100215809, 100215810, 100315078, 100315079, 100315080, 100315081, 100315082, 100315083, 100315084, 100315085, 100315086, 100315087, 100315088, 100315089, 100315090, 100315091, 100315092, 100315670, 100315671, 100315672, 100315673, 100315674, 100315675, 100315676, 100315677, 100315678, 100315679, 080815116, 091215376, 091215377, 091215378, 110215808, 110215809, 100815273, 100815274, 100815275, 100815276, 100815277, 100815278, 100815279, 100815280, 100815281, 100815282, 100815283, 100815284, 100815285, 100815286, 100815287, 100815288, 100815289, 100815290, 100815291, 100815292, 100815293, 100815294, 100815295, 101116017, 110215554, 100815726, 100815727, 100815728, 100815729, 100815730, 100815731, 100815732, 100815733, 100815734, 100815735, 100815736, 100815737, 100815738, 100815739, 100815740, 100815741, 100815742, 100815743, 100815744, 100815745, 100815746, 100815747, 100815748, 100815749, 100815750, 100815751, 100815752, 100815753, 100815754, 100815755, 100815756, 100815757, 100815758, 100815759, 100815760, 100815761, 100815884, 100815885, 100815886, 100815887, 100815888, 100815889, 100815890, 100815891, 100815892, 100815893, 100815894, 100815895, 100815896, 100815897, 100815898, 100815899, 100815900, 100815901, 100815902, 100815903, 100815904, 100815905, 100815906, 100815907, 100815908, 100815909, 100815910, 100815911, 100815912, 100815913, 100815914, 100815915, 100815916, 100815917, 100815918, 080815227, 090616274, 090616275, 090616276, 090616277, 090616278, 090616279, 080715001, 080715002, 080715003, 080715004, 080715005, 080715006, 080715007, 080715008, 080715009, 080715010, 080715011, 080715012, 080715013, 080715014, 080715015, 080715016, 080715017, 080715018, 080715019, 080715020, 080715021, 080715022, 080715023, 080715024, 080715025, 080715026, 080715027, 080715028, 080715029, 080715030, 080715031, 080715081, 080715082, 080715083, 080715084, 080715085, 080715086, 080715087, 080715088, 080715089, 080715090, 080715091, 080715092, 080715093, 080715094, 080715095, 080715096, 080715097, 080715098, 080715099, 080715100, 080715101, 080715102, 080715103, 080715104, 080715105, 080715106, 080715107, 080715108, 080715109, 080715110, 080715111, 080715114, 081215107, 090316108, 090415132, 090815025, 090815026, 090815027, 090815028, 090815029, 090815030, 090815031, 090815032, 090815033, 090815034, 090815035, 090815036, 090815037, 090815038, 090815039, 090815040, 090815041, 090815042, 090815043, 091116010, 091116011, 091116012, 091116013, 091116014, 091116015, 091116016, 091116017, 091116018, 091116019, 091116020, 091116021, 091116022, 091116023, 091116024, 091116025, 091116026, 091116027, 091116028, 091116029, 091116030, 091116031, 091116032, 091116033, 091116034, 091116035, 091116036, 091116037, 091116038, 091116039, 091116040, 100315027, 100315352, 100416123, 100515548, 100615297, 100615298, 100615300, 100615592, 100615696, 100615699, 100615700, 100615718, 100615829, 100616115, 100815876, 100815878, 100816127, 101015652, 101015653, 101015675, 101015677, 101015679, 101015955, 101015957, 101015958, 101015961, 101015962, 101016094, 101016097, 101016099, 101016100, 101016103, 10
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAProCuity bed series, model number 3009, item number: 3009PX-LE-400.December 13, 2023
- FDAProform- support surface is used to assist in the prevention and treatment of all categories/stages ofAugust 18, 2021
- FDAPrime Fifth Wheel Stretcher-stretcher may be used for minor procedures and short-term stay (treatment andAugust 18, 2021
- FDAThe Spirit Plus Bed, Model 5600 Spirit Plus /with high rail siderail optionNovember 21, 2018
- FDASecure II Med/Surg Bed Secure II and the 3002 S3 are the beds typically found in the Med-Surg units of theJanuary 8, 2014
- FDAAdvanDx Gram-Negative QuickFISH BC. In-Vitro Diagnostic for the identification of Enterobacteriacae CatalogJanuary 8, 2014
Questions about this recall
Was S3 Med/Surg Bed Secure II and the 3002 S3 are the beds typically found in the Med-Surg units of the hospitals. The bed is designed for a 15 year expected service life under normal use conditions, and with appropriate periodic maintenance as described in the maintenance manual for each device. There are minimal differences between the functionality of the two models, each of which function as a highly complex system of a large number of individual components. recalled?
Yes. FDA published recall Z-0618-2014 on January 8, 2014. Reported hazard: Stryker Medical has identified a potential issue associated with Secure II and S3 Med/Surg beds built between April 2nd, 2005 and July 26th, 2011. The affected beds have the potential to experience damage to the Foot End Cover and/or Power Coil Cable due to reduced clearance between the Foot End Cover and Foot End Lift Header.
What should I do if I have this product?
Voluntary: Firm initiated · Letter