PKG, AXIAL HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080235. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Published December 24, 2014 · Recall initiated: November 17, 2014
Summary
FDA records medical device Class II recall Z-0625-2015, published December 24, 2014 and initiated November 17, 2014. Recalling firm: Stryker Endoscopy; Status: Terminated; Reported quantity: 180,573 units total; Lot: 41811 140622 140726 140727 140729 342723 345252 544042 544043 644785 645845 645850 645852 645854 648085 06CON06065 744231 744232 745522 746745 840675 842159 842161 842197 842207 846841 846842 846844 846845 846847 846848 846849 846850 846851 846853 846855 942926 942943 942944 942945 942946 942947 942950 942951 942953 942954 942960 942964 943554 1044584 1044800 1045871 1046151 1046152 1046159 1046162 1046165 1140204 1140205 1140659 1140660 1140662 1142190 1142191 1143092 1143093 1143436 1143437 1144693 1144694 1144696 1145106 1145400 1145478 1146161 1146982 1240188 1240189 1241117 1241949 1242400 1243129 1243584 1243585 1244442 1244882 1246186 1246187 1340352 1340493 1340494 1341007 1341164 1341880 1342685 1342858 1344252 1344707 1440353 1442562 1443626 544043 645847. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Endoscopy
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lot
- 41811 140622 140726 140727 140729 342723 345252 544042 544043 644785 645845 645850 645852 645854 648085 06CON06065 744231 744232 745522 746745 840675 842159 842161 842197 842207 846841 846842 846844 846845 846847 846848 846849 846850 846851 846853 846855 942926 942943 942944 942945 942946 942947 942950 942951 942953 942954 942960 942964 943554 1044584 1044800 1045871 1046151 1046152 1046159 1046162 1046165 1140204 1140205 1140659 1140660 1140662 1142190 1142191 1143092 1143093 1143436 1143437 1144693 1144694 1144696 1145106 1145400 1145478 1146161 1146982 1240188 1240189 1241117 1241949 1242400 1243129 1243584 1243585 1244442 1244882 1246186 1246187 1340352 1340493 1340494 1341007 1341164 1341880 1342685 1342858 1344252 1344707 1440353 1442562 1443626 544043 645847
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,December 24, 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ROTATING HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080234. Laparoscopic Manual Instruments are intendedDecember 24, 2014
- FDAPKG, RING HANDLE, P/N 0250080236. Laparoscopic Manual Instruments are intended for cutting, holding, graspingDecember 24, 2014
Questions about this recall
Was PKG, AXIAL HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080235. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. recalled?
Yes. FDA published recall Z-0625-2015 on December 24, 2014. Reported hazard: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Voluntary: Firm initiated · Letter