FDAFDA · Food and drugs·Recall number Z-0628-2013
Medical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human patients. Lockout features may limit patient accessible controls
Published January 9, 2013 · Recall initiated: December 18, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. The affected beds may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Stryker Medical Division of Stryker Corporation
- Distributor
- Worldwide Distribution - Nationwide and Canada
- Lot
- Model 3002SEX, Serial Number 120315371-120315379
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States