Medical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human patients. Lockout features may limit patient accessible controls
Published January 9, 2013 · Recall initiated: December 18, 2012
Summary
FDA records medical device Class II recall Z-0628-2013, published January 9, 2013 and initiated December 18, 2012. Recalling firm: Stryker Medical Division of Stryker Corporation; Status: Terminated; Reported quantity: 9; Lot: Model 3002SEX, Serial Number 120315371-120315379. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. The affected beds may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Medical Division of Stryker Corporation
- Distributor
- Worldwide Distribution - Nationwide and Canada
- Lot
- Model 3002SEX, Serial Number 120315371-120315379
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAProCuity bed series, model number 3009, item number: 3009PX-LE-400.December 13, 2023
- FDAProform- support surface is used to assist in the prevention and treatment of all categories/stages ofAugust 18, 2021
- FDAPrime Fifth Wheel Stretcher-stretcher may be used for minor procedures and short-term stay (treatment andAugust 18, 2021
- FDAThe Spirit Plus Bed, Model 5600 Spirit Plus /with high rail siderail optionNovember 21, 2018
- FDAMaquet Yuno Extension Table Device Maquet GmbH & Co. KG Kehler Strabe 31 Baden-Wuerttemberg 76437January 9, 2013
- FDAMedical Surgical Bed, Model 3005S3EX Medical Surgical Bed The S3" MedSurg Bed, Model 3005 is intended toJanuary 9, 2013
Questions about this recall
Was Medical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human patients. Lockout features may limit patient accessible controls recalled?
Yes. FDA published recall Z-0628-2013 on January 9, 2013. Reported hazard: Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. The affected beds may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit