The Propaq LT Series monitors. Model numbers 802LTAN, 802LT0N, and 802LTRN. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medical-surgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. intended to be used by clinicians and medically qualified personnel for single or multiparameter vital signs monitoring of ambulatory and nonambulatory patients, including neonate, pediatric and adult patients. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medicalsurgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. The monitors can be used as stand-alone devices or as devices networked to an Acuity Central Station through wireless communication over a Welch Allyn FlexNet network.
Published January 15, 2014 · Recall initiated: November 21, 2013
Summary
FDA records medical device Class II recall Z-0633-2014, published January 15, 2014 and initiated November 21, 2013. Recalling firm: Welch Allyn Protocol, Inc; Status: Terminated; Reported quantity: 6003 units total (1715 units in the US and 4288 outside the US); Lot: SERIAL NUMBERS OF DEVICES DISTRIBUTED IN THE US: KA008788, KA009581, KA009685, KA009776, KA009917, KA010079, KA010193, KA010996, KA011141, KA011192, KA011227, KA011228, KA011231, KA011249, KA011303, KA011307, KA011311, KA011314, KA011327, KA011355, KA011356, KA011609, KA011610, KA011611, KA011667, KA011671, KA011672, KA011673, KA011674, KA011675, KA011676, KA011677, KA011678, KA011682, KA011731, KA011758, KA020006, KA020007, KA020008, KA020009, KA020010, KA020011, KA020012, KA020013, KA020014, KA020015, KA020016, KA020018, KA020023, KA020088, KA020091, KA020092, KA020093, KA020116, KA020135, KA020140, KA020149, KA020150, KA020151, KA020192, KA020200, KA020206, KA020207, KA020208, KA020209, KA020210, KA020211, KA020212, KA020213, KA020214, KA020215, KA020262, KA020264, KA020265, KA020266, KA020269, KA020270, KA020273, KA020295, KA020299, KA020304, KA020310, KA020311, KA020313, KA020314, KA020321, KA020381, KA020385, KA020386, KA020387, KA020388, KA020389, KA020390, KA020391, KA020392, KA020393, KA020394, KA020405, KA020406, KA020407, KA020408, KA020409, KA020410, KA020411, KA020412, KA020413, KA020414, KA020417, KA020422, KA020423, KA020424, KA020425, KA020426, KA020427, KA020429, KA020430, KA020432, KA020433, KA020435, KA020436, KA020437, KA020438, KA020439, KA020442, KA020443, KA020444, KA020445, KA020446, KA020447, KA020448, KA020449, KA020450, KA020451, KA020452, KA020453, KA020454, KA020455, KA020456, KA020457, KA020458, KA020459, KA020460, KA020461, KA020462, KA020463, KA020464, KA020465, KA020466, KA020467, KA020468, KA020469, KA020470, KA020471, KA020472, KA020473, KA020474, KA020475, KA020477, KA020478, KA020479, KA020480, KA020481, KA020482, KA020483, KA020526, KA020529, KA020532, KA020533, KA020534, KA020535, KA020536, KA020537, KA020538, KA020540, KA020541, KA020542, KA020543, KA020545, KA020546, KA020547, KA020548, KA020549, KA020550, KA020551, KA020552, KA020553, KA020554, KA020555, KA020556, KA020557, KA020558, KA020559, KA020560, KA020561, KA020562, KA020563, KA020564, KA020575, KA020576, KA020577, KA020578, KA020581, KA020582, KA020595, KA020596, KA020597, KA020606, KA020611, KA020613, KA020617, KA020631, KA020632, KA020633, KA020641, KA020646, KA020649, KA020650, KA020655, KA020659, KA020666, KA020685, KA020691, KA020692, KA020693, KA020726, KA020729, KA020742, KA020743, KA020746, KA020748, KA020853, KA020908, KA020909, KA020916, KA020917, KA021019, KA021028, KA021043, KA021048, KA021061, KA021062, KA021063, KA021064, KA021065, KA021086, KA021126, KA021151, KA021155, KA021158, KA021159, KA021170, KA021181, KA021182, KA021184, KA021188, KA021190, KA021202, KA021221, KA021277, KA021278, KA021279, KA021280, KA021286, KA021294, KA021298, KA021299, KA021311, KA021312, KA021313, KA021318, KA021319, KA021322, KA021323, KA021327, KA021330, KA021349, KA021365, KA021366, KA021367, KA021368, KA021369, KA021370, KA021371, KA021392, KA021393, KA021394, KA021395, KA021396, KA021397, KA021398, KA021399, KA021401, KA021402, KA021403, KA021406, KA021407, KA021408, KA021410, KA021413, KA021415, KA021417, KA021419, KA021426, KA021427, KA021429, KA021462, KA021465, KA021466, KA021469, KA021486, KA021492, KA021495, KA021503, KA021511, KA021516, KA021522, KA021524, KA021529, KA021531, KA021534, KA021545, KA021546, KA021547, KA021549, KA021552, KA021553, KA021556, KA021557, KA021558, KA021559, KA021560, KA021562, KA021563, KA021564, KA021565, KA021566, KA021567, KA021568, KA021569, KA021570, KA021572, KA021576, KA021587, KA021588, KA021589, KA021590, KA021591, KA021592, KA021598, KA021606, KA021609, KA021611, KA021613, KA021614, KA021615, KA021616, KA021623, KA021624, KA021626, KA021630, KA021632, KA021635, KA021670, KA021679, KA021702, KA021704, KA021705, KA021706, KA021707, KA021708, KA021709, KA021711, KA021713, KA021714, KA021715, KA021716, KA021722, KA021723, KA021741, KA021742, KA021745, KA021746, KA021748, KA021749, KA021750, KA021765, KA021766, KA021786, KA021787, KA021788, KA021790, KA021799, KA021803, KA021804, KA021805. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Welch Allyn will update Propaq 802 Series Vital Signs Monitor, models are 802LTAN, 802LTON and 802LTR with software because the fault conditions relating to hardware or software failure or electrical noise caused by proximal defibrillation activity could, in rare instances, potentially interrupt device operations.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Welch Allyn Protocol, Inc
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to United Arab Emirates, Argentina, Australia, Belgium, Brazil, Canada, Switzerland, Chile, China, Columbia, Costa Rica, Czech Republic, Germany, Denmark, Ecuador, Spain, France, United Kingdom, Greece, Hong Kong,Taiwan, Ireland, Italy, Japan, Korea, Republic of, Kuwait, Lithuania, Luxemborg, Mexico, Malaysia, Mozambique, Namibia, Netherlands, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Sierra Leone, El Salvador, Thailand, Turkey, Uganda, Vietnam, Kosovo, and South Africa.
- Lot
- SERIAL NUMBERS OF DEVICES DISTRIBUTED IN THE US: KA008788, KA009581, KA009685, KA009776, KA009917, KA010079, KA010193, KA010996, KA011141, KA011192, KA011227, KA011228, KA011231, KA011249, KA011303, KA011307, KA011311, KA011314, KA011327, KA011355, KA011356, KA011609, KA011610, KA011611, KA011667, KA011671, KA011672, KA011673, KA011674, KA011675, KA011676, KA011677, KA011678, KA011682, KA011731, KA011758, KA020006, KA020007, KA020008, KA020009, KA020010, KA020011, KA020012, KA020013, KA020014, KA020015, KA020016, KA020018, KA020023, KA020088, KA020091, KA020092, KA020093, KA020116, KA020135, KA020140, KA020149, KA020150, KA020151, KA020192, KA020200, KA020206, KA020207, KA020208, KA020209, KA020210, KA020211, KA020212, KA020213, KA020214, KA020215, KA020262, KA020264, KA020265, KA020266, KA020269, KA020270, KA020273, KA020295, KA020299, KA020304, KA020310, KA020311, KA020313, KA020314, KA020321, KA020381, KA020385, KA020386, KA020387, KA020388, KA020389, KA020390, KA020391, KA020392, KA020393, KA020394, KA020405, KA020406, KA020407, KA020408, KA020409, KA020410, KA020411, KA020412, KA020413, KA020414, KA020417, KA020422, KA020423, KA020424, KA020425, KA020426, KA020427, KA020429, KA020430, KA020432, KA020433, KA020435, KA020436, KA020437, KA020438, KA020439, KA020442, KA020443, KA020444, KA020445, KA020446, KA020447, KA020448, KA020449, KA020450, KA020451, KA020452, KA020453, KA020454, KA020455, KA020456, KA020457, KA020458, KA020459, KA020460, KA020461, KA020462, KA020463, KA020464, KA020465, KA020466, KA020467, KA020468, KA020469, KA020470, KA020471, KA020472, KA020473, KA020474, KA020475, KA020477, KA020478, KA020479, KA020480, KA020481, KA020482, KA020483, KA020526, KA020529, KA020532, KA020533, KA020534, KA020535, KA020536, KA020537, KA020538, KA020540, KA020541, KA020542, KA020543, KA020545, KA020546, KA020547, KA020548, KA020549, KA020550, KA020551, KA020552, KA020553, KA020554, KA020555, KA020556, KA020557, KA020558, KA020559, KA020560, KA020561, KA020562, KA020563, KA020564, KA020575, KA020576, KA020577, KA020578, KA020581, KA020582, KA020595, KA020596, KA020597, KA020606, KA020611, KA020613, KA020617, KA020631, KA020632, KA020633, KA020641, KA020646, KA020649, KA020650, KA020655, KA020659, KA020666, KA020685, KA020691, KA020692, KA020693, KA020726, KA020729, KA020742, KA020743, KA020746, KA020748, KA020853, KA020908, KA020909, KA020916, KA020917, KA021019, KA021028, KA021043, KA021048, KA021061, KA021062, KA021063, KA021064, KA021065, KA021086, KA021126, KA021151, KA021155, KA021158, KA021159, KA021170, KA021181, KA021182, KA021184, KA021188, KA021190, KA021202, KA021221, KA021277, KA021278, KA021279, KA021280, KA021286, KA021294, KA021298, KA021299, KA021311, KA021312, KA021313, KA021318, KA021319, KA021322, KA021323, KA021327, KA021330, KA021349, KA021365, KA021366, KA021367, KA021368, KA021369, KA021370, KA021371, KA021392, KA021393, KA021394, KA021395, KA021396, KA021397, KA021398, KA021399, KA021401, KA021402, KA021403, KA021406, KA021407, KA021408, KA021410, KA021413, KA021415, KA021417, KA021419, KA021426, KA021427, KA021429, KA021462, KA021465, KA021466, KA021469, KA021486, KA021492, KA021495, KA021503, KA021511, KA021516, KA021522, KA021524, KA021529, KA021531, KA021534, KA021545, KA021546, KA021547, KA021549, KA021552, KA021553, KA021556, KA021557, KA021558, KA021559, KA021560, KA021562, KA021563, KA021564, KA021565, KA021566, KA021567, KA021568, KA021569, KA021570, KA021572, KA021576, KA021587, KA021588, KA021589, KA021590, KA021591, KA021592, KA021598, KA021606, KA021609, KA021611, KA021613, KA021614, KA021615, KA021616, KA021623, KA021624, KA021626, KA021630, KA021632, KA021635, KA021670, KA021679, KA021702, KA021704, KA021705, KA021706, KA021707, KA021708, KA021709, KA021711, KA021713, KA021714, KA021715, KA021716, KA021722, KA021723, KA021741, KA021742, KA021745, KA021746, KA021748, KA021749, KA021750, KA021765, KA021766, KA021786, KA021787, KA021788, KA021790, KA021799, KA021803, KA021804, KA021805
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAnspach B-Blue-S 11.25 cm attachment and Anspach B-Green Standard Craniotome. Cutting and shaping bone.January 15, 2014
- FDATracer IV wheelchairs, Heavy duty manual wheel chair, 24" seat width, 450 lbs. weight capacity. Distributed byJanuary 15, 2014
- FDAAnspach BLACKMAX-N (Pneumatic motor) and Autolube III (Foot Control). Cutting and shaping bone.January 15, 2014
- FDASandel Healthcare Safety Devices Weighted Safety Scalpel #3 Disposable Safety Handle with #10 BladeJanuary 15, 2014
- FDAStryker Color Cuff, REF 5921-034-135 Disposable Tourniquet Cuffs 34 in x 4 in (86 cm x 10 cm) Single Bladder,January 15, 2014
- FDASandel Healthcare Safety Devices Weighted Safety Scalpel #3 Disposable Safety Handle with #11 BladeJanuary 15, 2014
Questions about this recall
Was The Propaq LT Series monitors. Model numbers 802LTAN, 802LT0N, and 802LTRN. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medical-surgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. intended to be used by clinicians and medically qualified personnel for single or multiparameter vital signs monitoring of ambulatory and nonambulatory patients, including neonate, pediatric and adult patients. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medicalsurgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. The monitors can be used as stand-alone devices or as devices networked to an Acuity Central Station through wireless communication over a Welch Allyn FlexNet network. recalled?
Yes. FDA published recall Z-0633-2014 on January 15, 2014. Reported hazard: Welch Allyn will update Propaq 802 Series Vital Signs Monitor, models are 802LTAN, 802LTON and 802LTR with software because the fault conditions relating to hardware or software failure or electrical noise caused by proximal defibrillation activity could, in rare instances, potentially interrupt device operations.
What should I do if I have this product?
Voluntary: Firm initiated · Letter