Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT/2INJ/FLBL/FML LRLK/CV 34; Catalog Number: V5405.

Published December 3, 2025 · Recall initiated: October 29, 2025

Summary

FDA records medical device Class II recall Z-0639-2026, published December 3, 2025 and initiated October 29, 2025. Recalling firm: B Braun Medical Inc; Status: Ongoing; Reported quantity: 300 units; Lot: 1. Catalog Number: V5405; Primary UDI-DI: 04046964188978; Unit of Dose UDI-DI: 04046964188961. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
B Braun Medical Inc
Distributor
Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Lot
1. Catalog Number: V5405; Primary UDI-DI: 04046964188978; Unit of Dose UDI-DI: 04046964188961.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT/2INJ/FLBL/FML LRLK/CV 34; Catalog Number: V5405. recalled?

Yes. FDA published recall Z-0639-2026 on December 3, 2025. Reported hazard: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

What should I do if I have this product?

Voluntary: Firm initiated · Letter