Incisive CT, model 728143 & 728144 running Software Version 5.0.0.
Published January 17, 2024 · Recall initiated: December 23, 2022
Summary
FDA records medical device Class II recall Z-0640-2024, published January 17, 2024 and initiated December 23, 2022. Recalling firm: Philips Healthcare (Suzhou) Co., Ltd.; Status: Ongoing; Reported quantity: 828; Lot: UDI-DI: 00884838085015 (Model 728143) UDI-DI: 00884838105508 (Model 728144). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan. These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operators voice are too quiet, patient cant hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Healthcare (Suzhou) Co., Ltd.
- Distributor
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, RI, SC, TN, TX, UT, VT, WI, & Puerto Rico and the countries of Albania, Algeria, Argentina, Australia, Austria, Bhutan, Bolivia, Bosnia-Herz, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Mauritius, Moldova, Morocco, Netherlands, New Zealand, Norway, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Senegal, Serbia, Sint Maarten, Slovakia, South Africa, South Korea, Spain, St. Pier Miquel, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, United Arab Emirates, Uzbekistan, Venezuela, & Vietnam.
- Lot
- UDI-DI: 00884838085015 (Model 728143) UDI-DI: 00884838105508 (Model 728144)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- China
Related recalls
- FDAPhilips Incisive CT with software V4.5&V5.0, Product Code 728143 and Product Code 728144.September 2, 2026
- FDAPhilips Incisive CTJanuary 14, 2026
- FDABrand Name: SUPREX Plant Based Nutrition. Product Name: Carb & Sugar Block. Product Size: 120 VegetableJanuary 17, 2024
- FDACD11b APC: ASR, REF: 340936, and CE, REF: 333143January 17, 2024
- FDA3/4" Diced Tomatoes in Juice packed in 55 gallon metal drumsJanuary 17, 2024
- FDABard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/January 17, 2024
Questions about this recall
Was Incisive CT, model 728143 & 728144 running Software Version 5.0.0. recalled?
Yes. FDA published recall Z-0640-2024 on January 17, 2024. Reported hazard: Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan. These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operators voice are too quiet, patient cant hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP.
What should I do if I have this product?
Voluntary: Firm initiated · Letter