Back to search
FDAFDA · Food and drugs·Recall number Z-0640-2024

Incisive CT, model 728143 & 728144 running Software Version 5.0.0.

Published January 17, 2024 · Recall initiated: December 23, 2022

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan. These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operators voice are too quiet, patient cant hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips Healthcare (Suzhou) Co., Ltd.
Distributor
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, RI, SC, TN, TX, UT, VT, WI, & Puerto Rico and the countries of Albania, Algeria, Argentina, Australia, Austria, Bhutan, Bolivia, Bosnia-Herz, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Mauritius, Moldova, Morocco, Netherlands, New Zealand, Norway, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Senegal, Serbia, Sint Maarten, Slovakia, South Africa, South Korea, Spain, St. Pier Miquel, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, United Arab Emirates, Uzbekistan, Venezuela, & Vietnam.
Lot
UDI-DI: 00884838085015 (Model 728143) UDI-DI: 00884838105508 (Model 728144)
Risk classification
Class II
Area (FDA)
device
Country
China

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card