ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100
Published January 24, 2024 · Recall initiated: December 7, 2023
Summary
FDA records medical device Class I recall Z-0642-2024, published January 24, 2024 and initiated December 7, 2023. Recalling firm: Globus Medical, Inc.; Status: Ongoing; Reported quantity: 9; Lot: UDI-DI 00193982465155 Serial Numbers FPFF-0101 FPFF-0102 FPFF-0103 FPFF-0104 FPFF-0105 FPFF-0106 FPFF-0107 FPFF-0108 FPFF-0109. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Globus Medical, Inc.
- Distributor
- Domestic distribution to the following states: CA, CO, MT, SC, WA. International distribution to Germany and UK.
- Lot
- UDI-DI 00193982465155 Serial Numbers FPFF-0101 FPFF-0102 FPFF-0103 FPFF-0104 FPFF-0105 FPFF-0106 FPFF-0107 FPFF-0108 FPFF-0109
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGlobus Medical Cranial Drill, 6.00mm - instruments for access, preparation, and placement of cranial devicesFebruary 14, 2024
- FDAGlobus Medical Cranial Drill with Stop, 3.20mm - instruments for access, preparation, and placement of cranialFebruary 14, 2024
- FDAGlobus Medical Cranial Drill with Stop, 2.40mm - Instruments for access, preparation, and placement of cranialFebruary 14, 2024
- FDAExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844February 14, 2024
- FDAHigh Frequency Transport Phasitron Breathing Circuit Kit REF: A50605-D, Rx Only, CE 0123January 24, 2024
- FDADistal Phasitron, Part: S20020, for use with Percussionaire VDR-4 control driver (or control unit)January 24, 2024
Questions about this recall
Was ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100 recalled?
Yes. FDA published recall Z-0642-2024 on January 24, 2024. Reported hazard: A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.
What should I do if I have this product?
Voluntary: Firm initiated · Letter