ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100

Published January 24, 2024 · Recall initiated: December 7, 2023

Summary

FDA records medical device Class I recall Z-0642-2024, published January 24, 2024 and initiated December 7, 2023. Recalling firm: Globus Medical, Inc.; Status: Ongoing; Reported quantity: 9; Lot: UDI-DI 00193982465155 Serial Numbers FPFF-0101 FPFF-0102 FPFF-0103 FPFF-0104 FPFF-0105 FPFF-0106 FPFF-0107 FPFF-0108 FPFF-0109. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Globus Medical, Inc.
Distributor
Domestic distribution to the following states: CA, CO, MT, SC, WA. International distribution to Germany and UK.
Lot
UDI-DI 00193982465155 Serial Numbers FPFF-0101 FPFF-0102 FPFF-0103 FPFF-0104 FPFF-0105 FPFF-0106 FPFF-0107 FPFF-0108 FPFF-0109
Risk classification
Class I
Area (FDA)
device
Country
United States