PKG, METZENBAUM SCISSORS, CURVED LONG, P/N 0250080265. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Published December 24, 2014 · Recall initiated: November 17, 2014
Summary
FDA records medical device Class II recall Z-0644-2015, published December 24, 2014 and initiated November 17, 2014. Recalling firm: Stryker Endoscopy; Status: Terminated; Reported quantity: 180,573 units; Lot: 140280 143823 244698 341493 443333 541776 545061 647416 744587 842092 844289 846109 846772 846819 941064 941065 941852 942140 943501 943549 944300 945744 945886 945977 1040098 1040413 1041057 1041621 1041992 1042086 1042240 1042475 1042685 1042695 1042698 1042714 1044482 1045035 1045373 1045641 1046199 1046595 1046933 1046934 1046943 1142155 1142668 1142897 1142910 1143132 1144208 1144398 1144492 1144940 1145080 1145798 1145971 1146839 1146933 1240119 1241514 1241515 1242921 1244191 1244534 1245445 1246192E 1246638 1246638E 1246650 1340099 1341584 1341585 1343125 1343402 1343402E 1343705 1344204 1344286 1344456 1344880 1345236 1345532 1345532E 1345605 1345605E 1346432 1346666 1440405 1440684 1442518 1443219 1443653 1443917. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Endoscopy
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lot
- 140280 143823 244698 341493 443333 541776 545061 647416 744587 842092 844289 846109 846772 846819 941064 941065 941852 942140 943501 943549 944300 945744 945886 945977 1040098 1040413 1041057 1041621 1041992 1042086 1042240 1042475 1042685 1042695 1042698 1042714 1044482 1045035 1045373 1045641 1046199 1046595 1046933 1046934 1046943 1142155 1142668 1142897 1142910 1143132 1144208 1144398 1144492 1144940 1145080 1145798 1145971 1146839 1146933 1240119 1241514 1241515 1242921 1244191 1244534 1245445 1246192E 1246638 1246638E 1246650 1340099 1341584 1341585 1343125 1343402 1343402E 1343705 1344204 1344286 1344456 1344880 1345236 1345532 1345532E 1345605 1345605E 1346432 1346666 1440405 1440684 1442518 1443219 1443653 1443917
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,December 24, 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, METZENBAUM SCISSORS, STRAIGHT, P/N 0250080264. Laparoscopic Manual Instruments are intended for cutting,December 24, 2014
- FDAPKG, MINI METZENBAUM SCISSORS, CURVED LONG, P/N 0250080266. Laparoscopic Manual Instruments are intended forDecember 24, 2014
Questions about this recall
Was PKG, METZENBAUM SCISSORS, CURVED LONG, P/N 0250080265. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. recalled?
Yes. FDA published recall Z-0644-2015 on December 24, 2014. Reported hazard: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Voluntary: Firm initiated · Letter