Sterile Rigid Tip Suction Wand The Suction Wands are indicated for use to remove excess fluid from the surgical field.
Published January 16, 2013 · Recall initiated: November 2, 2012
Summary
FDA records medical device Class II recall Z-0649-2013, published January 16, 2013 and initiated November 2, 2012. Recalling firm: California Medical Laboratories Inc; Status: Terminated; Reported quantity: S112462=995 units & S111480 1430 units; Lot: Catalog No: SU-12202 Lot No: S112462, S111480. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
California Medical Laboratories, Inc. has identified an issue with the Rigid Tip Suction Wand. Due to insufficient adhesive application, the distal tip has the potential to come loose from the wand and fall into the patient's chest cavity.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- California Medical Laboratories Inc
- Distributor
- Worldwide Distribution - USA including AZ, FL, CA, LA, CO, and MS. Internationally to Canada, Israel, Greece, Hong Kong, India, Ireland, Jordan, Lebanon, Malaysia, Netherlands, Russia, Saudi Arabia, South Africa and Spain.
- Lot
- Catalog No: SU-12202 Lot No: S112462, S111480
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARigid Tip Suction Wand Non sterile bulk devices (S112468N, S120196N, and S120197N) were sold to Terumo UK toJanuary 16, 2013
- FDAImmunoradiometric assay. Device is an immunoradiometric assay designed for the quantitative measurement ofJanuary 16, 2013
- FDAImmunoradiometric assay.January 16, 2013
- FDAUltrafill DBM labeled in part: TISSUENET Sterilized by T10^6 Description: UltraFill DBM Size: 1cc ProductJanuary 16, 2013
- FDAImmunoradiometric assay. Device is an immunoradiometric assay designed for the quantitative measurement ofJanuary 16, 2013
- FDAUltraFill DBM labeled in part: TISSUENET Sterilized by T10^6 Description: UltraFill DBM Size: 5cc ProductJanuary 16, 2013
Questions about this recall
Was Sterile Rigid Tip Suction Wand The Suction Wands are indicated for use to remove excess fluid from the surgical field. recalled?
Yes. FDA published recall Z-0649-2013 on January 16, 2013. Reported hazard: California Medical Laboratories, Inc. has identified an issue with the Rigid Tip Suction Wand. Due to insufficient adhesive application, the distal tip has the potential to come loose from the wand and fall into the patient's chest cavity.
What should I do if I have this product?
Voluntary: Firm initiated · Letter