Back to search
FDAFDA · Food and drugs·Recall number Z-0654-2021

Star 8 Rolling Walker/Rollator, Items: 4258/ BK(Black)/ BL(Blue)/ DB(Sky Blue)/ PL(Purple)/ RD(Red)

Published December 30, 2020 · Recall initiated: October 7, 2019

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Walkers within the identified serial range have the potential for the front fork to unscrew itself from the leg. A partially unscrewed fork creates unevenness in the wheelbase, which could cause instability, that if not detected by the user, may result in the front fork falling off the walker.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Nova Ortho-Med Inc
Distributor
U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, WY O.U.S.: Taiwan
Lot
Serial # range: HF-4258BKHA0001 to HF4258RDHK0352 (Purchased before April 4th,2019)
Risk classification
Class II
Area (FDA)
device
Country
United States