5000 Series Freedom Drivers. Freedom Driver System for Temporary Total Artificial Heart (TAH-t). Part number 595000-001

Published December 14, 2016 · Recall initiated: October 21, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The Main Printed Circuit Board Assembly (PCBA) of the affected Freedom Drivers may fail and cause the Freedom Driver to stop functioning without visual or audible alarms, resulting in the loss of life-sustaining function.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
SynCardia Systems Inc.
Distributor
US distribution to Virginia and Arizona.
Lot
S/N 5002  UDI (01)00858000003121(21)5002 S/N 5005  UDI (01)00858000003121(21)5005 S/N 5008  UDI (01)00858000003121(21)5008 S/N 5009  UDI (01)00858000003121(21)5009 S/N 5010  UDI (01)00858000003121(21)5010
Risk classification
Class I
Area (FDA)
device
Country
United States