MICRO-X Rover Mobile X-ray System, # MXU-RV35

Published December 18, 2024 · Recall initiated: May 31, 2023

Summary

FDA records medical device Class II recall Z-0664-2025, published December 18, 2024 and initiated May 31, 2023. Recalling firm: Micro-X Ltd.; Status: Ongoing; Reported quantity: 5; Lot: UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Micro-X Ltd.
Distributor
Worldwide and US Nationwide Distribution: NJ, PR, CA
Lot
UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿
Risk classification
Class II
Area (FDA)
device
Country
Australia