MICRO-X Rover Mobile X-ray System, # MXU-RV35
Published December 18, 2024 · Recall initiated: May 31, 2023
Summary
FDA records medical device Class II recall Z-0664-2025, published December 18, 2024 and initiated May 31, 2023. Recalling firm: Micro-X Ltd.; Status: Ongoing; Reported quantity: 5; Lot: UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Micro-X Ltd.
- Distributor
- Worldwide and US Nationwide Distribution: NJ, PR, CA
- Lot
- UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Australia
Related recalls
- FDAMICRO-X Rover Mobile X-ray System, # MXU-RV19December 18, 2024
- FDAMICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71December 18, 2024
- FDAMicroscissor Tip (Disposable); Microline Catalog #: 3122.December 18, 2024
- FDAChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer ofDecember 18, 2024
- FDAMICRO-X Rover Mobile X-ray System, # MXU-RV35December 4, 2024
- FDAMICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19October 5, 2022
Questions about this recall
Was MICRO-X Rover Mobile X-ray System, # MXU-RV35 recalled?
Yes. FDA published recall Z-0664-2025 on December 18, 2024. Reported hazard: Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
What should I do if I have this product?
Voluntary: Firm initiated · Letter