Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus. ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus.

Published January 17, 2024 · Recall initiated: November 30, 2023

Summary

FDA records medical device Class II recall Z-0673-2024, published January 17, 2024 and initiated November 30, 2023. Recalling firm: Boston Scientific Corporation; Status: Ongoing; Reported quantity: 14,328 ZOOM programmers; Lot: All serial numbers. GTIN numbers in the U.S.: 00802526434723, 00802526456961, 00802526467653, 00802526468582, 00802526473449, 00802526474859, 00802526475788, 00802526477676, 00802526478239, 00802526479663, 00802526480638, 00802526484599, 00802526486906, 00802526494406, 00802526497605, 00802526516009, 00802526523212, 00802526535406, 00802526554209, 00802526562600, 00802526564703, 00802526575303, 00802526579707, 00802526582004, 00802526602900, 00802526604584, 00802526606366, 00802526606403, and 00802526611322. GTIN numbers OUS: 00802526435430, 00802526468582, 00802526479663, 00802526480638, 00802526486906, 00802526478246, 00802526479670, 00802526484605, 00802526486913, 00802526496608, 00802526496615, 00802526506703, 00802526515903, 00802526517112, 00802526526602, 00802526527609, 00802526554803, 00802526561900, 00802526564802, 00802526574801, 00802526579905, 00802526582011, 00802526602917, 00802526605406, 00802526606342, 00802526611308, 00802526615603, 00802526437458, 00802526441974, 00802526444814, 00802526467288, 00802526469343, 00802526475450, 00802526475757, 00802526477669, 00802526478253, 00802526479687, 00802526484612, 00802526486920, 00802526496622, 00802526496639, 00802526515910, 00802526517136, 00802526526800, 00802526527616, 00802526554810, 00802526564819, 00802526574818, 00802526581021, 00802526582028, 00802526605437, 00802526606311, 00802526612428, 00802526615634, 00802526454301, 00802526467295, 00802526469350, 00802526475467, 00802526475764, 00802526477676, 00802526484803, 00802526486081, 00802526486937, 00802526497629, 00802526515927, 00802526532832, 00802526557002, 00802526565427, 00802526582301, 00802526587405, 00802526602924, 00802526606359, 00802526612442, 00802526445262, 00802526454950, 00802526474170, 00802526475474, 00802526475771, 00802526480652, 00802526486944, 00802526497643, 00802526517174, 00802526526817, 00802526527623, 00802526557019, 00802526565304, 00802526579929, 00802526583803, 00802526602931, 00802526605468, 00802526606304, 00802526612411, 00802526615627, 00802526442827, 00802526444807, 00802526469374, 00802526477683, 00802526482267, 00802526497650, 00802526517198, 00802526540400, 00802526555169, 00802526564826, 00802526582325, 00802526585104, 00802526586316, 00802526586316, 00802526603013, 00802526606328, 00802526612435, 00802526615429, 00802526474057, 00802526477522, 00802526479649, 00802526482274, 00802526486470, 00802526497667, 00802526517211, 00802526540509, 00802526555206, 00802526564833, 00802526582318, 00802526585111, 00802526586408, 00802526603006, 00802526605420, 00802526606335, 00802526612404, 00802526615504, 00802526480669, 00802526605451, 00802526496653, 00802526526824, 00802526527630, 00802526527630, 00802526564840, 00802526581038, 00802526582035, 00802526584602, 00802526582035, 00802526605451, 00802526606403, and 00802526612459. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
Distribution was nationwide, including Guam, Puerto Rico, and the U.S. Virgin Islands. There was also government/military distribution. Foreign distribution was made to Andorra, Argentina, Aruba, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, British Virgin Islands, Bulgaria, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Faroe Islands, Finland, France, French Polynesia, Germany, Great Britain, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Yemen, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan R.O.C., Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, Uruguay, Venezuela, Vietnam, and Zimbabwe.
Lot
All serial numbers. GTIN numbers in the U.S.: 00802526434723, 00802526456961, 00802526467653, 00802526468582, 00802526473449, 00802526474859, 00802526475788, 00802526477676, 00802526478239, 00802526479663, 00802526480638, 00802526484599, 00802526486906, 00802526494406, 00802526497605, 00802526516009, 00802526523212, 00802526535406, 00802526554209, 00802526562600, 00802526564703, 00802526575303, 00802526579707, 00802526582004, 00802526602900, 00802526604584, 00802526606366, 00802526606403, and 00802526611322. GTIN numbers OUS: 00802526435430, 00802526468582, 00802526479663, 00802526480638, 00802526486906, 00802526478246, 00802526479670, 00802526484605, 00802526486913, 00802526496608, 00802526496615, 00802526506703, 00802526515903, 00802526517112, 00802526526602, 00802526527609, 00802526554803, 00802526561900, 00802526564802, 00802526574801, 00802526579905, 00802526582011, 00802526602917, 00802526605406, 00802526606342, 00802526611308, 00802526615603, 00802526437458, 00802526441974, 00802526444814, 00802526467288, 00802526469343, 00802526475450, 00802526475757, 00802526477669, 00802526478253, 00802526479687, 00802526484612, 00802526486920, 00802526496622, 00802526496639, 00802526515910, 00802526517136, 00802526526800, 00802526527616, 00802526554810, 00802526564819, 00802526574818, 00802526581021, 00802526582028, 00802526605437, 00802526606311, 00802526612428, 00802526615634, 00802526454301, 00802526467295, 00802526469350, 00802526475467, 00802526475764, 00802526477676, 00802526484803, 00802526486081, 00802526486937, 00802526497629, 00802526515927, 00802526532832, 00802526557002, 00802526565427, 00802526582301, 00802526587405, 00802526602924, 00802526606359, 00802526612442, 00802526445262, 00802526454950, 00802526474170, 00802526475474, 00802526475771, 00802526480652, 00802526486944, 00802526497643, 00802526517174, 00802526526817, 00802526527623, 00802526557019, 00802526565304, 00802526579929, 00802526583803, 00802526602931, 00802526605468, 00802526606304, 00802526612411, 00802526615627, 00802526442827, 00802526444807, 00802526469374, 00802526477683, 00802526482267, 00802526497650, 00802526517198, 00802526540400, 00802526555169, 00802526564826, 00802526582325, 00802526585104, 00802526586316, 00802526586316, 00802526603013, 00802526606328, 00802526612435, 00802526615429, 00802526474057, 00802526477522, 00802526479649, 00802526482274, 00802526486470, 00802526497667, 00802526517211, 00802526540509, 00802526555206, 00802526564833, 00802526582318, 00802526585111, 00802526586408, 00802526603006, 00802526605420, 00802526606335, 00802526612404, 00802526615504, 00802526480669, 00802526605451, 00802526496653, 00802526526824, 00802526527630, 00802526527630, 00802526564840, 00802526581038, 00802526582035, 00802526584602, 00802526582035, 00802526605451, 00802526606403, and 00802526612459.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus. ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus. recalled?