DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.
Published January 23, 2013 · Recall initiated: July 24, 2012
Summary
FDA records medical device Class II recall Z-0685-2013, published January 23, 2013 and initiated July 24, 2012. Recalling firm: DePuy Spine, Inc.; Status: Terminated; Reported quantity: 140 units; Lot: All lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DePuy Spine, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK.
- Lot
- All lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPIPELINE LS ACCESS SYSTEM, Holding Clip The PIPELINE LS Dilators and Holding Clip are intended for sequentialNovember 6, 2013
- FDAPIPELINE LS ACCESS SYSTEM, 3rd Dilator, 14 mm Insulated The PIPELINE LS Dilators and Holding Clip are intendedNovember 6, 2013
- FDAPIPELINE LS ACCESS SYSTEM, 15 Dilator, 5 mm, Insulated The PIPLINE LS Dilators and Holding Clip are intendedNovember 6, 2013
- FDAPIPELINE LS ACCESS SYSTEM, 2nd Dilator, 10 mm, Insulated The PIPELINE LS Dilators and Holding Clip areNovember 6, 2013
- FDAFresenius GranuFlo Naturalyte Dry Acid Concentrate For Bicarbonate Dialysis 3.0K 2,5 Ca 0.75Mg .16.5GALJanuary 23, 2013
- FDANEOGRAM AMINO ACIDS AND ACYLCARNITINES TANDEM MASS SPECTROMETRY KIT Product Number: 3026-0030. This kit isJanuary 23, 2013
Questions about this recall
Was DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct. recalled?
Yes. FDA published recall Z-0685-2013 on January 23, 2013. Reported hazard: A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.
What should I do if I have this product?
Voluntary: Firm initiated · Letter