FDAFDA · Food and drugs·Recall number Z-0685-2013
DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.
Published January 23, 2013 · Recall initiated: July 24, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- DePuy Spine, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK.
- Lot
- All lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States