DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.

Published January 23, 2013 · Recall initiated: July 24, 2012

Summary

FDA records medical device Class II recall Z-0685-2013, published January 23, 2013 and initiated July 24, 2012. Recalling firm: DePuy Spine, Inc.; Status: Terminated; Reported quantity: 140 units; Lot: All lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
DePuy Spine, Inc.
Distributor
Worldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK.
Lot
All lots
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct. recalled?

Yes. FDA published recall Z-0685-2013 on January 23, 2013. Reported hazard: A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.

What should I do if I have this product?

Voluntary: Firm initiated · Letter