Model No. BF-XP190, EVIS EXERA III Bronchovideoscope
Published January 17, 2024 · Recall initiated: December 4, 2023
Summary
FDA records medical device Class II recall Z-0687-2024, published January 17, 2024 and initiated December 4, 2023. Recalling firm: Shirakawa Olympus Co., Ltd.; Status: Ongoing; Reported quantity: 1187 US; 1652 OUS; Lot: UDI-DI: 04953170342134 All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Shirakawa Olympus Co., Ltd.
- Distributor
- US Nationwide distribution.
- Lot
- UDI-DI: 04953170342134 All serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Japan
Related recalls
- FDAModel No. CYF-VH, CYSTO-NEPHRO VIDEOSCOPEJanuary 17, 2024
- FDAModel No. CYF-VHR, CYSTO-NEPHRO VIDEOSCOPEJanuary 17, 2024
- FDAModel No. CYF-V2R, VISERA Cysto-Nephro VideoscopeJanuary 17, 2024
- FDAModel No. CYF-5R, CystonephrofiberscopeJanuary 17, 2024
- FDAModel No. BF-P190, EVIS EXERA III BronchovideoscopeJanuary 17, 2024
- FDAModel No. ENF-XP, RHINO-LARYNGOFIBERSCOPEJanuary 17, 2024
Questions about this recall
Was Model No. BF-XP190, EVIS EXERA III Bronchovideoscope recalled?
Yes. FDA published recall Z-0687-2024 on January 17, 2024. Reported hazard: Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
What should I do if I have this product?
Voluntary: Firm initiated