PKG, NEEDLE HOLDER, CURVED RIGHT, P/N 0250080342. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Published December 24, 2014 · Recall initiated: November 17, 2014
Summary
FDA records medical device Class II recall Z-0691-2015, published December 24, 2014 and initiated November 17, 2014. Recalling firm: Stryker Endoscopy; Status: Terminated; Reported quantity: 180,573 units total; Lot: 344320 542914 842447 843440 844144 844173 844321 844957 845072 846763 941831 942655 945439 1040398 1040425 1043259 1044647 1046103 1140613 1140855 1141922 1142930 1144227 1144948 1241790 1241791 1241945 1242479 1244068 1245027 1246068 1246193 1340477 1341248 1342749 1343416 1343711 1344464 1440436 1441385. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Endoscopy
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lot
- 344320 542914 842447 843440 844144 844173 844321 844957 845072 846763 941831 942655 945439 1040398 1040425 1043259 1044647 1046103 1140613 1140855 1141922 1142930 1144227 1144948 1241790 1241791 1241945 1242479 1244068 1245027 1246068 1246193 1340477 1341248 1342749 1343416 1343711 1344464 1440436 1441385
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,December 24, 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, NEEDLE HOLDER, CURVED LEFT, P/N 0250080341. Laparoscopic Manual Instruments are intended for cutting,December 24, 2014
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 45CM, P/N 0250080350. Laparoscopic Manual Instruments are intended forDecember 24, 2014
Questions about this recall
Was PKG, NEEDLE HOLDER, CURVED RIGHT, P/N 0250080342. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. recalled?
Yes. FDA published recall Z-0691-2015 on December 24, 2014. Reported hazard: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Voluntary: Firm initiated · Letter