Back to search
FDAFDA · Food and drugs·Recall number Z-0693-2015

PKG, BIPOLAR FORCEPS, PADDLE, 45CM, P/N 0250080351. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Published December 24, 2014 · Recall initiated: November 17, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Stryker Endoscopy
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lot
816247 816248 817088 817415 817476 913562 913563 913828 913829 913830 915433 915678 915679 915680 915792 915891 915892 915893 915949 1013276 1013838 1014368 1014638 1014639 1014750 1014943 1015257 1016064 1016065 1016066 1016067 1016198 1016250 1016735 1016737 1016738 1214971 1215297 1215300 1216025 1216026 1216027 1216028 1216822 1216823 1216824 1217033 1315269 1315270 1315932 1315985 1316201 1316305 1410091 1410847 1410875 1913567
Risk classification
Class II
Area (FDA)
device
Country
United States