Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Purple Cap, Bag Hanger, REF: 011-CH3185; 46 cm (18") Appx 7.4 ml, Trifuse Add-On Set w/2 Spiros w/Purple Caps, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: 011-CH3266; 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Purple Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: 011-CH3446; 61 cm (24") Appx 8.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros w/Purple Cap, 3 Clamps (1 Blue, 2 White), Check Valve, Dry Spike Adapter, REF: 011-CH3729; 25 UNITS of Spiros CLOSED MALE LUER, REF: 011-H2457; Appx 1.0 ml, Adattatore per Pompa con Spiros, REF: 011-H2629; 179 cm (70") Appx 11.5 ml, PUR Ambr. Deflussore con Spiros, Camera, BCV, Y-Clave e LL Gir., REF: 011-H2776; 180 cm (71IN)APPX 11.5ml,PUR AMBR,Spiros,CAMERA,REG.FLUSSO,BCV,Y-Clave, REF: 011-H2864-HSR; 181cm(71 IN)APPX11.6ml,PUR DEFLUSSORE,Spiros,20 DROP IN LINE,BCV,Clave, REF: 011-H2904; 38 cm (15 IN) APPX 2.9 ml, TRIFUSE EXT SET, 3 MicroClave, Spiros, PURPLE CAP, REF: 011-H3449; 42cm(16.5 IN)APPX 2.6ml,PUR TRANSFER SET,Clave ADDITIVE PORT,BCV,Spiros, REF: 011-H3597; 41cm(16 IN)APPX 2.9ml,PUR SET PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3628; 41cm(16 IN)APPX 3.1ml,SET AMBRATO,PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3629; 42cm(16.5 IN)APPX 5.6ml,SET,PERFORATORE,Clave,FILTRO,VALVOLA,Spiros, REF: 011-H3630; 109 cm(43 IN) APPX 7.2 ml,KIT,EXT SET W/Spiros,STOPCOCK,5 Clave,4-GANG STOPCOCK, REF: 011-MC330632; 2 Ext Sets w/5-Gang Stopcock, 9 Clave, Spiros, Stopcock, REF: 011-MC33242; Kit for multiple transfers, REF: 011-N1000; Kit per il riempimento pompe elastomeriche con SPIROS / Kit for elastomeric pump priming with SPIROS, REF: 011-N1000-P2; APPX 2.3ml, ZYTO-VERBINDUNGSLEITUNG (ZYTO-CONNECTING LINE), REF: 011-N1664SP; Appx 1.0 ml, Adaptador universal para sistemas IV con Spiros, REF: 034-H2629; Spinning Spiros Closed Male Luer, Red Cap (No Flange), REF: CH2000S-NR; 15" (38 cm) Appx 5.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros, 3 Clamps (Blue, 2 White), Dry Spike Adapter, REF: CH3187; 14" (36 cm) Appx 8.0 ml, Quinfuse Add-On Set w/4 Spiros, Bag Spike, 5 Clamps (Blue, 4 Red), Dry Spike Adapter, REF: CH3225; 20" (51 cm) Appx 9.5 ml, Quadfuse Add-On Set w/3 Spiros, Bag Spike, 4 Clamps (Blue, 3 Red), Dry Spike Adapter, REF: CH3226; 16" (41 cm) Appx 6.3 ml, Trifuse Add-On Set w/2 Spiros, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: CH3227; 30" (76 cm) Appx 4.1 ml, Yellow 20 Drop Admin Set w/15 Micron Filter, Spiros, REF: CH3235; 28" (71 cm) Appx 5.1 ml, Add On Set w/Clave Spike, Spiros, Clamp, REF: CH3354; 16cm(6 IN)APPX 2.4ml,SET,MicroClave Clear,STOPCOCK,DROP-IN ADD-ON,Spiros, REF CH3411: APPX 1.9 ml,13 cm (5 IN),BAG SPIKE,Spiros,ACCESS DEVICE w/Clave, REF: CH3502; 60 (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; Spinning Spiros Closed Male Luer, REF: K7040-001; APPX 0.76ml, 4-WAY STOPCOCK w/2 Clave, VIAL ADAPTER, 3 Spiros, REF: K7052-001; ZYTOmed Verab.-Set Onko II lichtschutz, REF: SN1205AG; ZYTOmed - Verabreichungs-Set Onko III lichtgesch®tzt, REF: SN1206A; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC
Published December 30, 2020 · Recall initiated: August 31, 2020
Summary
FDA records medical device Class II recall Z-0706-2021, published December 30, 2020 and initiated August 31, 2020. Recalling firm: ICU Medical, Inc.; Status: Ongoing; Reported quantity: 217522; Lot: Lots: 4630846, 4848415, 4734764, 4734940, 4752044, 4742983, 4764957, 4878172, 4808952, 4867502, 4804800, 4854732, 4783273, 4773067, 4807223, 4800168, 4877026, 4752842, 4773432, 4764194, 4727034, 4847780, 4876626, 4548846, 4573171, 4867154, 4515064, 4574113, 4743010, 4773247, 4848995, 4740817, 4774162, 4877106, 4902461, 4711498, 4719930, 4740835, 4798726, 4866446, 4894490, 4727444, 4741918, 4778013, 4800251, 4806881, 4743840, 4734281, 4775528, 4799761, 4763261, 4774080, 4790739, 4749875, 4794599, 4802672, 4859281, 4718351, 4727042, 4741435, 4774258, 4866217, 4875126, 4899256, 4774269, 4777289, 4849437, 4853031, 4847272, 4882644 ,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ICU Medical, Inc.
- Distributor
- Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
- Lot
- Lots: 4630846, 4848415, 4734764, 4734940, 4752044, 4742983, 4764957, 4878172, 4808952, 4867502, 4804800, 4854732, 4783273, 4773067, 4807223, 4800168, 4877026, 4752842, 4773432, 4764194, 4727034, 4847780, 4876626, 4548846, 4573171, 4867154, 4515064, 4574113, 4743010, 4773247, 4848995, 4740817, 4774162, 4877106, 4902461, 4711498, 4719930, 4740835, 4798726, 4866446, 4894490, 4727444, 4741918, 4778013, 4800251, 4806881, 4743840, 4734281, 4775528, 4799761, 4763261, 4774080, 4790739, 4749875, 4794599, 4802672, 4859281, 4718351, 4727042, 4741435, 4774258, 4866217, 4875126, 4899256, 4774269, 4777289, 4849437, 4853031, 4847272, 4882644 ,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAOncology sets with drip chambersSeptember 23, 2026
- FDAIV Non-Dedicated Gravity Administration Sets with drip chambersSeptember 23, 2026
- FDA30" (76 cm) Appx 5.5 mL 20 Drop Admin Set w/Integrated Clave Drip Chamber, 1.2 Micron Filter, Spiros, BagDecember 30, 2020
- FDA26 IN(66 cm) APPX 3.4ml, 10 DROP BLOOD SET, 170 MICRON FILTER, Spiros, REF: Z2795December 30, 2020
Questions about this recall
Was Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Purple Cap, Bag Hanger, REF: 011-CH3185; 46 cm (18") Appx 7.4 ml, Trifuse Add-On Set w/2 Spiros w/Purple Caps, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: 011-CH3266; 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Purple Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: 011-CH3446; 61 cm (24") Appx 8.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros w/Purple Cap, 3 Clamps (1 Blue, 2 White), Check Valve, Dry Spike Adapter, REF: 011-CH3729; 25 UNITS of Spiros CLOSED MALE LUER, REF: 011-H2457; Appx 1.0 ml, Adattatore per Pompa con Spiros, REF: 011-H2629; 179 cm (70") Appx 11.5 ml, PUR Ambr. Deflussore con Spiros, Camera, BCV, Y-Clave e LL Gir., REF: 011-H2776; 180 cm (71IN)APPX 11.5ml,PUR AMBR,Spiros,CAMERA,REG.FLUSSO,BCV,Y-Clave, REF: 011-H2864-HSR; 181cm(71 IN)APPX11.6ml,PUR DEFLUSSORE,Spiros,20 DROP IN LINE,BCV,Clave, REF: 011-H2904; 38 cm (15 IN) APPX 2.9 ml, TRIFUSE EXT SET, 3 MicroClave, Spiros, PURPLE CAP, REF: 011-H3449; 42cm(16.5 IN)APPX 2.6ml,PUR TRANSFER SET,Clave ADDITIVE PORT,BCV,Spiros, REF: 011-H3597; 41cm(16 IN)APPX 2.9ml,PUR SET PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3628; 41cm(16 IN)APPX 3.1ml,SET AMBRATO,PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3629; 42cm(16.5 IN)APPX 5.6ml,SET,PERFORATORE,Clave,FILTRO,VALVOLA,Spiros, REF: 011-H3630; 109 cm(43 IN) APPX 7.2 ml,KIT,EXT SET W/Spiros,STOPCOCK,5 Clave,4-GANG STOPCOCK, REF: 011-MC330632; 2 Ext Sets w/5-Gang Stopcock, 9 Clave, Spiros, Stopcock, REF: 011-MC33242; Kit for multiple transfers, REF: 011-N1000; Kit per il riempimento pompe elastomeriche con SPIROS / Kit for elastomeric pump priming with SPIROS, REF: 011-N1000-P2; APPX 2.3ml, ZYTO-VERBINDUNGSLEITUNG (ZYTO-CONNECTING LINE), REF: 011-N1664SP; Appx 1.0 ml, Adaptador universal para sistemas IV con Spiros, REF: 034-H2629; Spinning Spiros Closed Male Luer, Red Cap (No Flange), REF: CH2000S-NR; 15" (38 cm) Appx 5.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros, 3 Clamps (Blue, 2 White), Dry Spike Adapter, REF: CH3187; 14" (36 cm) Appx 8.0 ml, Quinfuse Add-On Set w/4 Spiros, Bag Spike, 5 Clamps (Blue, 4 Red), Dry Spike Adapter, REF: CH3225; 20" (51 cm) Appx 9.5 ml, Quadfuse Add-On Set w/3 Spiros, Bag Spike, 4 Clamps (Blue, 3 Red), Dry Spike Adapter, REF: CH3226; 16" (41 cm) Appx 6.3 ml, Trifuse Add-On Set w/2 Spiros, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: CH3227; 30" (76 cm) Appx 4.1 ml, Yellow 20 Drop Admin Set w/15 Micron Filter, Spiros, REF: CH3235; 28" (71 cm) Appx 5.1 ml, Add On Set w/Clave Spike, Spiros, Clamp, REF: CH3354; 16cm(6 IN)APPX 2.4ml,SET,MicroClave Clear,STOPCOCK,DROP-IN ADD-ON,Spiros, REF CH3411: APPX 1.9 ml,13 cm (5 IN),BAG SPIKE,Spiros,ACCESS DEVICE w/Clave, REF: CH3502; 60 (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; Spinning Spiros Closed Male Luer, REF: K7040-001; APPX 0.76ml, 4-WAY STOPCOCK w/2 Clave, VIAL ADAPTER, 3 Spiros, REF: K7052-001; ZYTOmed Verab.-Set Onko II lichtschutz, REF: SN1205AG; ZYTOmed - Verabreichungs-Set Onko III lichtgesch®tzt, REF: SN1206A; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC recalled?
Yes. FDA published recall Z-0706-2021 on December 30, 2020. Reported hazard: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
What should I do if I have this product?
Voluntary: Firm initiated · Letter