OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.

Published December 25, 2024 · Recall initiated: November 21, 2024

Summary

FDA records medical device Class II recall Z-0706-2025, published December 25, 2024 and initiated November 21, 2024. Recalling firm: GE OEC Medical Systems, Inc; Status: Ongoing; Reported quantity: 25 systems; Lot: Product Name: OEC 9800 GTIN code: 00840682114349 Lot/Serial Numbers / UDI codes: 82XXXX04237 (01)00840682114349(11)170600(21)82XXXX04237 8S-2249 Not Applicable-MFG date prior to UDI compliance requirement 89-1022 Not Applicable-MFG date prior to UDI compliance requirement 82-1772 Not Applicable-MFG date prior to UDI compliance requirement 82-0980 Not Applicable-MFG date prior to UDI compliance requirement 8S-1793 Not Applicable-MFG date prior to UDI compliance requirement 82-2349 Not Applicable-MFG date prior to UDI compliance requirement 82-3361 Not Applicable-MFG date prior to UDI compliance requirement 8S-1398 Not Applicable-MFG date prior to UDI compliance requirement 82-0908 Not Applicable-MFG date prior to UDI compliance requirement 82-0599 Not Applicable-MFG date prior to UDI compliance requirement 8S-0533 Not Applicable-MFG date prior to UDI compliance requirement 82-0189 Not Applicable-MFG date prior to UDI compliance requirement 8S-0186 Not Applicable-MFG date prior to UDI compliance requirement 89-2415 Not Applicable-MFG date prior to UDI compliance requirement 8S-3259 Not Applicable-MFG date prior to UDI compliance requirement 8S-3426 Not Applicable-MFG date prior to UDI compliance requirement 89-1530 Not Applicable-MFG date prior to UDI compliance requirement 82-2955 Not Applicable-MFG date prior to UDI compliance requirement 8S-0958-RC Not Applicable-MFG date prior to UDI compliance requirement 82-1154 Not Applicable-MFG date prior to UDI compliance requirement 8S-3032 Not Applicable-MFG date prior to UDI compliance requirement 8S-3145 Not Applicable-MFG date prior to UDI compliance requirement 89-1916-RC Not Applicable-MFG date prior to UDI compliance requirement 82-4134 Not Applicable-MFG date prior to UDI compliance requirement. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE OEC Medical Systems, Inc
Distributor
U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom
Lot
Product Name: OEC 9800 GTIN code: 00840682114349 Lot/Serial Numbers / UDI codes: 82XXXX04237 (01)00840682114349(11)170600(21)82XXXX04237 8S-2249 Not Applicable-MFG date prior to UDI compliance requirement 89-1022 Not Applicable-MFG date prior to UDI compliance requirement 82-1772 Not Applicable-MFG date prior to UDI compliance requirement 82-0980 Not Applicable-MFG date prior to UDI compliance requirement 8S-1793 Not Applicable-MFG date prior to UDI compliance requirement 82-2349 Not Applicable-MFG date prior to UDI compliance requirement 82-3361 Not Applicable-MFG date prior to UDI compliance requirement 8S-1398 Not Applicable-MFG date prior to UDI compliance requirement 82-0908 Not Applicable-MFG date prior to UDI compliance requirement 82-0599 Not Applicable-MFG date prior to UDI compliance requirement 8S-0533 Not Applicable-MFG date prior to UDI compliance requirement 82-0189 Not Applicable-MFG date prior to UDI compliance requirement 8S-0186 Not Applicable-MFG date prior to UDI compliance requirement 89-2415 Not Applicable-MFG date prior to UDI compliance requirement 8S-3259 Not Applicable-MFG date prior to UDI compliance requirement 8S-3426 Not Applicable-MFG date prior to UDI compliance requirement 89-1530 Not Applicable-MFG date prior to UDI compliance requirement 82-2955 Not Applicable-MFG date prior to UDI compliance requirement 8S-0958-RC Not Applicable-MFG date prior to UDI compliance requirement 82-1154 Not Applicable-MFG date prior to UDI compliance requirement 8S-3032 Not Applicable-MFG date prior to UDI compliance requirement 8S-3145 Not Applicable-MFG date prior to UDI compliance requirement 89-1916-RC Not Applicable-MFG date prior to UDI compliance requirement 82-4134 Not Applicable-MFG date prior to UDI compliance requirement
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was OEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. recalled?

Yes. FDA published recall Z-0706-2025 on December 25, 2024. Reported hazard: Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

What should I do if I have this product?

Voluntary: Firm initiated · Letter