DePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component.

Published January 30, 2013 · Recall initiated: November 7, 2012

Summary

FDA records medical device Class II recall Z-0712-2013, published January 30, 2013 and initiated November 7, 2012. Recalling firm: DePuy Spine, Inc.; Status: Terminated; Reported quantity: 6 units; Lot: All lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
DePuy Spine, Inc.
Distributor
Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.
Lot
All lots
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component. recalled?

Yes. FDA published recall Z-0712-2013 on January 30, 2013. Reported hazard: Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head

What should I do if I have this product?

Voluntary: Firm initiated · Letter