DePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component.
Published January 30, 2013 · Recall initiated: November 7, 2012
Summary
FDA records medical device Class II recall Z-0712-2013, published January 30, 2013 and initiated November 7, 2012. Recalling firm: DePuy Spine, Inc.; Status: Terminated; Reported quantity: 6 units; Lot: All lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- DePuy Spine, Inc.
- Distributor
- Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.
- Lot
- All lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPIPELINE LS ACCESS SYSTEM, Holding Clip The PIPELINE LS Dilators and Holding Clip are intended for sequentialNovember 6, 2013
- FDAPIPELINE LS ACCESS SYSTEM, 3rd Dilator, 14 mm Insulated The PIPELINE LS Dilators and Holding Clip are intendedNovember 6, 2013
- FDAPIPELINE LS ACCESS SYSTEM, 15 Dilator, 5 mm, Insulated The PIPLINE LS Dilators and Holding Clip are intendedNovember 6, 2013
- FDAPIPELINE LS ACCESS SYSTEM, 2nd Dilator, 10 mm, Insulated The PIPELINE LS Dilators and Holding Clip areNovember 6, 2013
- FDADePuy Spine VIPER POLY SCREW 5.5 10 X 70MM Tl Product Code: 1867-15-170 Spinal Implant Component.January 30, 2013
- FDADePuy Spine VIPER POLY SCREW 5.5 10X100MM Tl Product Code: 1867-15-199 Spinal Implant Component.January 30, 2013
Questions about this recall
Was DePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component. recalled?
Yes. FDA published recall Z-0712-2013 on January 30, 2013. Reported hazard: Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head
What should I do if I have this product?
Voluntary: Firm initiated · Letter