Locking Cortical Screw 3.5mmX38mm REF 8161-35-038
Published January 22, 2014 · Recall initiated: December 19, 2013
Summary
FDA records medical device Class II recall Z-0716-2014, published January 22, 2014 and initiated December 19, 2013. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 52; Lot: Catalog Number: 816135038 (8161-35-038) Lot Number Identification: RM244D. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
3.5mmx38mm Locking Cortical Screw, implants are being recalled since an investigation identified that a 3.5mmx28mm Cortical Locking Screw is labeled as a 3.5mmx38mm Cortical Locking Screw.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- US Distribution including states of: AZ, CA, GA, IL, KY, MI, NC, NH, NY, OH, PA, SD, TX, UT, and VA.
- Lot
- Catalog Number: 816135038 (8161-35-038) Lot Number Identification: RM244D
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475625 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary BoneAugust 12, 2026
- FDAAnspach MCA Cutting Burr, DFU Rx Only Designed for cutting and shaping delicate bone, primarily in otologyJanuary 22, 2014
- FDALeica BIOSYSTEMS; BOND Polymer Refine Red Detection; Catalog No: DS9390; Leica Biosystems Newcastle Ltd;January 22, 2014
Questions about this recall
Was Locking Cortical Screw 3.5mmX38mm REF 8161-35-038 recalled?
Yes. FDA published recall Z-0716-2014 on January 22, 2014. Reported hazard: 3.5mmx38mm Locking Cortical Screw, implants are being recalled since an investigation identified that a 3.5mmx28mm Cortical Locking Screw is labeled as a 3.5mmx38mm Cortical Locking Screw.
What should I do if I have this product?
Voluntary: Firm initiated · Letter