PKG, ALLIS DOUBLE ROW TEETH, P/N 0250080592 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Published December 24, 2014 · Recall initiated: November 17, 2014
Summary
FDA records medical device Class II recall Z-0717-2015, published December 24, 2014 and initiated November 17, 2014. Recalling firm: Stryker Endoscopy; Status: Terminated; Reported quantity: 180,573 units total; Lot: 140289 641421 840358 848483 940108 940131 946000 1040117 1040987 1041626 1042737 1042745 1045058 1046928 1143580 1145635 1145811 1145930 1146540 1241586 1242142 1242643 1243443 1243554 1244523 1340562 1341469 1341471 1341995 1343407 1343706 1344781 1346104 1346430 1346651 1346652. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Endoscopy
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lot
- 140289 641421 840358 848483 940108 940131 946000 1040117 1040987 1041626 1042737 1042745 1045058 1046928 1143580 1145635 1145811 1145930 1146540 1241586 1242142 1242643 1243443 1243554 1244523 1340562 1341469 1341471 1341995 1343407 1343706 1344781 1346104 1346430 1346651 1346652
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,December 24, 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, 5MM INSERT, ALLIS SINGLE ROW TEETH, P/N 0250080591 Laparoscopic Manual Instruments are intended forDecember 24, 2014
- FDAPKG, 5MM ROTATING HANDLE, INSULATED SHAFT/ HANDLE 45CM LENGTH, P/N 0250080595 Laparoscopic Manual InstrumentsDecember 24, 2014
Questions about this recall
Was PKG, ALLIS DOUBLE ROW TEETH, P/N 0250080592 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. recalled?
Yes. FDA published recall Z-0717-2015 on December 24, 2014. Reported hazard: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
What should I do if I have this product?
Voluntary: Firm initiated · Letter