Max Implant System - Bone Tap for use with 7.0mm diameter Max Implant, Catalog Number D-TAP-MAX7-7 Dental implants are intended for use as anchors of fixed or semi-fixed dental crowns, bridges, and overdentures in patients with partial or full edentulism of the upper or lower jaw.
Published March 7, 2018 · Recall initiated: June 16, 2017
Summary
FDA records medical device Class II recall Z-0718-2018, published March 7, 2018 and initiated June 16, 2017. Recalling firm: Keystone Dental Inc; Status: Terminated; Reported quantity: 72; Lot: Catalog Number D-TAP-MAX7-7, Lot 31444. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An 8mm diameter bone tap was laser marked as 7mm diameter and was packed with a 7mm label.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Keystone Dental Inc
- Distributor
- USA (nationwide) Distribution to the states of : CA, CO, FL, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NH, NY, OH, PA, TN, TX, UT, VA, VT, WA, WI and WV; and Internationally to Canada
- Lot
- Catalog Number D-TAP-MAX7-7, Lot 31444
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPaltop Fully Guided Surgical Kit- containing the 1.25mm Hex Driver For surgical and restorative applicationsApril 3, 2019
- FDAPaltop Premium Surgical Kit-containing the 1.25mm Hex Driver For surgical and restorative applicationsApril 3, 2019
- FDAPrima Plus ¿4.1 x 11.5mm Implant, Catalog Number 15737K Keystone Dental implants are intended for use inJune 20, 2018
- FDAAllura Xper systems with Certeray X-ray Generator located in regions with high mains (480V) voltage. ProductMarch 7, 2018
- FDAultraView Universal DAB Detection Kit, Ventana part Number 760-500, Roche GMMI 05269806001 in vitro diagnosticMarch 7, 2018
- FDAPrimaConnex Tapered Implant TC, SD Internal Connection 3.5 x 10mm Catalog Number: l5613K. Implants in supportMarch 27, 2013
Questions about this recall
Was Max Implant System - Bone Tap for use with 7.0mm diameter Max Implant, Catalog Number D-TAP-MAX7-7 Dental implants are intended for use as anchors of fixed or semi-fixed dental crowns, bridges, and overdentures in patients with partial or full edentulism of the upper or lower jaw. recalled?
Yes. FDA published recall Z-0718-2018 on March 7, 2018. Reported hazard: An 8mm diameter bone tap was laser marked as 7mm diameter and was packed with a 7mm label.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail