Halyard C-Section Drape with Clear Screen and Full Incise (Non-Sterile), Product Code 44965 NS; C-Section Drape with Clear Screen and Full Incise (Sterile), Product Code 44966 00; C-Section Drape with Clear Screen (Non-Sterile), Product Code 44967 NS; C-Section Drape with Clear Screen (Sterile), Product Code 44968 00; C-Section Drape with Pouch and Fenestration (Non-Sterile), Product Code 44977 NS; C-Section Drape with Pouch and Fenestration (Sterile), Product Code 44978 00 Product Usage: A surgical drape is a device made of natural or synthetic materials intended to be use as a protective patient covering, such as to isolate a site or surgical incision from microbial and other contamination.
Published December 14, 2016 · Recall initiated: November 9, 2016
Summary
FDA records medical device Class II recall Z-0724-2017, published December 14, 2016 and initiated November 9, 2016. Recalling firm: Halyard Health, Inc; Status: Terminated; Reported quantity: 132 cases; Lot: Product Code 44965 NS - Lot Numbers AC6174###, AC62266###, AC6229###, AC6243###, AC6252###; Product Code 44966 00 - Lot Numbers AC6146##L, AC6157##L, AC6177##L, AC6215##L, AC6221##L, AC6228##L, AC6245##L, AC6252##L, AC6258##L, AC6272##L, AC6277##L; Product Code 44967 NS - Lot Numbers AC6170###, AC6219###, AC6223###, AC6229###, AC625###; Product Code 44068 00 - Lot Numbers AC6146##L, AC6177##L, AC6230##L, AC6231##L, AC6234##L, AC6242##L, AC6245##L, AC6264##L, AC6271##L, AC6277##L; Product Code 44077 NS - Lot Numbers AC6170###, AC6213###, AC6214###, AC6222###, AC6223###, AC6251###; Product Code 44978 00 - Lot Numbers AC6147##L, AC6177##L, AC6243##L, AC 6252##L - (# = includes numbers 0, 1, 2, 3, 4, 5, 6, 7, 8 or 9 - L = includes letters a, b, or c). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Drapes within the affected lots may include a manufacturing variation that prevents convenient removal of the coated release paper, which inhibits easy access to the surgical site.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Halyard Health, Inc
- Distributor
- US Nationwide Distribution in the states of AL, FL, GA, IL, KS, MN, MO, NE, NJ, OH, PH, TN, VA, WI, WY
- Lot
- Product Code 44965 NS - Lot Numbers AC6174###, AC62266###, AC6229###, AC6243###, AC6252###; Product Code 44966 00 - Lot Numbers AC6146##L, AC6157##L, AC6177##L, AC6215##L, AC6221##L, AC6228##L, AC6245##L, AC6252##L, AC6258##L, AC6272##L, AC6277##L; Product Code 44967 NS - Lot Numbers AC6170###, AC6219###, AC6223###, AC6229###, AC625###; Product Code 44068 00 - Lot Numbers AC6146##L, AC6177##L, AC6230##L, AC6231##L, AC6234##L, AC6242##L, AC6245##L, AC6264##L, AC6271##L, AC6277##L; Product Code 44077 NS - Lot Numbers AC6170###, AC6213###, AC6214###, AC6222###, AC6223###, AC6251###; Product Code 44978 00 - Lot Numbers AC6147##L, AC6177##L, AC6243##L, AC 6252##L - (# = includes numbers 0, 1, 2, 3, 4, 5, 6, 7, 8 or 9 - L = includes letters a, b, or c)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHalyard Closed Suction System for Adults with BALLARD Technology WET PAK. 12 saline vials per kit, 20 kits perNovember 29, 2017
- FDAHalyard Closed Suction System for Adults, 14 F, DSE, Wet Pak. 12 or 6 vials per kit, 20 kits per case.November 29, 2017
- FDAHalyard Closed Suction System for Adults, 14 F, T-Piece, Wet Pak. 12 or 6 vials per kit, 20 kits per case.November 29, 2017
- FDAHalyard Turbo-Cleaning Closed Suction System for Adults, 14 F, T-Piece, Wet Pak. 12 vials per kit, 20 kits perNovember 29, 2017
- FDAADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD.December 14, 2016
- FDAMagnesium on RX Imola analyser IVDDecember 14, 2016
Questions about this recall
Was Halyard C-Section Drape with Clear Screen and Full Incise (Non-Sterile), Product Code 44965 NS; C-Section Drape with Clear Screen and Full Incise (Sterile), Product Code 44966 00; C-Section Drape with Clear Screen (Non-Sterile), Product Code 44967 NS; C-Section Drape with Clear Screen (Sterile), Product Code 44968 00; C-Section Drape with Pouch and Fenestration (Non-Sterile), Product Code 44977 NS; C-Section Drape with Pouch and Fenestration (Sterile), Product Code 44978 00 Product Usage: A surgical drape is a device made of natural or synthetic materials intended to be use as a protective patient covering, such as to isolate a site or surgical incision from microbial and other contamination. recalled?
Yes. FDA published recall Z-0724-2017 on December 14, 2016. Reported hazard: Drapes within the affected lots may include a manufacturing variation that prevents convenient removal of the coated release paper, which inhibits easy access to the surgical site.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit