Back to search
FDAFDA · Food and drugs·Recall number Z-0728-2013

Electrosurgical Tip Cleaner, REF/Catalog No. 138029, Single Use, Sterile, Rx ONLY -- Distributed By: CONMED CORPORATION Electrosurgical, Cutting and Coagulation and Accessories.

Published January 30, 2013 · Recall initiated: December 27, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

ConMed received complaints of some devices which had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products. In no instance has it been reported to ConMed that a compromise in the sterile barrier has resulted in illness or injury. The compromised seals were discovered prior to use.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
ConMed Corporation
Distributor
Worldwide Distribution - USA (nationwide) Puerto Rico and internationally to Belgium, Colombia, Czech Republic, Dubai, Guatemala, Hong Kong, India, Israel, Italy, Japan, Kenya, Kuwait, Malaysia, Mexico, New Zealand, Norway, Oman, Pakistan, Portugal, Saudi Arabia, Thailand, and Uruguay.
Lot
Beginning Lot Code 120628X, Ending Lot Code 120725X. [Lot codes on boxes and packaging contain a lot code in the following form: 120725X where 12 represents the year (2012), 07 represents the month (July), 25 represents the day of the month, and X represents the manufacturing shift code (1, 2, 3, 4,5,6 or 9)].
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card