VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
Published December 28, 2022 · Recall initiated: November 3, 2022
Summary
FDA records medical device Class II recall Z-0728-2023, published December 28, 2022 and initiated November 3, 2022. Recalling firm: Icotec Ag; Status: Ongoing; Reported quantity: 6 devices; Lot: UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Icotec Ag
- Distributor
- US Nationwide distribution in the state of GA.
- Lot
- UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Switzerland
Related recalls
- FDAVADER pedicle system, torque wrench, Catalog Number 42-702January 17, 2024
- FDAVADER pedicle system, torque wrench, Catalog Number 42-703January 17, 2024
- FDADenticator PICK-A-DENT, periodontal aid - intradental pick, REF 621714December 28, 2022
- FDAFull Dose CE Simplex P Single Bone Cement - Indicated for the fixation of prostheses to living bone inDecember 28, 2022
- FDAConvaTec AQUACEL Foam Ag Adhesive, wound dressing advesiveDecember 28, 2022
- FDAHeartWare Model 1100December 28, 2022
Questions about this recall
Was VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702 recalled?
Yes. FDA published recall Z-0728-2023 on December 28, 2022. Reported hazard: During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail