Back to search
FDAFDA · Food and drugs·Recall number Z-0728-2023

VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702

Published December 28, 2022 · Recall initiated: November 3, 2022

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Icotec Ag
Distributor
US Nationwide distribution in the state of GA.
Lot
UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045
Risk classification
Class II
Area (FDA)
device
Country
Switzerland

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card