VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702

Published December 28, 2022 · Recall initiated: November 3, 2022

Summary

FDA records medical device Class II recall Z-0728-2023, published December 28, 2022 and initiated November 3, 2022. Recalling firm: Icotec Ag; Status: Ongoing; Reported quantity: 6 devices; Lot: UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Icotec Ag
Distributor
US Nationwide distribution in the state of GA.
Lot
UDI/DI: (01) 07640164844589, Serial Numbers: 200221-040, 200221-041, 200221-042, 200221-043, 200221-044, 200221-045
Risk classification
Class II
Area (FDA)
device
Country
Switzerland

Related recalls

Questions about this recall

Was VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702 recalled?

Yes. FDA published recall Z-0728-2023 on December 28, 2022. Reported hazard: During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring results for both instruments were 4 Nm.