FDAFDA · Food and drugs·Recall number Z-0741-2013

HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics

Published February 6, 2013 · Recall initiated: December 19, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Hamilton Medical Inc
Distributor
US Nationwide Distribution including the country of Switzerland.
Lot
Software Versions 1.1.2 and below; Part numbers 161005, 161006.
Risk classification
Class I
Area (FDA)
device
Country
United States