FDAFDA · Food and drugs·Recall number Z-0741-2013

HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics

Published February 6, 2013 · Recall initiated: December 19, 2012

Summary

FDA records medical device Class I recall Z-0741-2013, published February 6, 2013 and initiated December 19, 2012. Recalling firm: Hamilton Medical Inc; Status: Terminated; Reported quantity: 97 total ventilators; 69 of part number 161005 and 28 of part number 161006; Lot: Software Versions 1.1.2 and below; Part numbers 161005, 161006. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Hamilton Medical Inc
Distributor
US Nationwide Distribution including the country of Switzerland.
Lot
Software Versions 1.1.2 and below; Part numbers 161005, 161006.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics recalled?

Yes. FDA published recall Z-0741-2013 on February 6, 2013. Reported hazard: Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.

What should I do if I have this product?

Voluntary: Firm initiated · Letter