Back to search
FDAFDA · Food and drugs·Recall number Z-0745-2017

BD 23 G x .75 in. BD Vacutainer¿ Push Button Blood Collection Set with 12 in. Tubing Catalog # 367342 Intended for venipuncture to obtain blood specimens from patients.

Published December 21, 2016 · Recall initiated: August 8, 2016

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

BD has received multiple reports of units associated with the 23 G x .75 in. BD Vacutainer¿ Push Button Blood Collection Set with 12 in. Tubing whose lots are exhibiting separation of front and rear barrels upon activation of the safety feature, which retracts the needle.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Becton Dickinson & Company
Distributor
Worldwide distribution- US (nationwide) and countries of: Canada, Belgium, Brazil.
Lot
5355531 5355533 5356681 5356682 6006629 6006630 6006631 6006633 6006634 6006650
Risk classification
Class II
Area (FDA)
device
Country
United States