Extremity Medical, IO FiX System Accessory. Catalog Number 101-00010. This screw measuring device is an accessory to the IO FiX Screw and Washer System that is used to measure the length of the screw by placing the depth gauge over the guidewire and down to the bone. Orthopedic device.
Published February 6, 2013 · Recall initiated: March 22, 2012
Summary
FDA records medical device Class II recall Z-0752-2013, published February 6, 2013 and initiated March 22, 2012. Recalling firm: Extremity Medical LLC; Status: Terminated; Reported quantity: 93 units; Lot: Catalog Number 101-00010 Lot number AP3V12 AP4E01 AP4N04 AP5E25 AP6A44 AP6E42 AP6L14 AP6M17 AP6T08. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Lag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical Technique Rev C. which can lead to possible inappropriate implant size selection (e.g. shorter screw size selected).
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Extremity Medical LLC
- Distributor
- Worldwide Distribution -- Nationwide Distribution including the states of AR, AZ, CA, FL, GA, IL, IN, MD, MT, NC, NE, NH, NJ, NY, OH, PA, TN, TX, WI and WA.
- Lot
- Catalog Number 101-00010 Lot number AP3V12 AP4E01 AP4N04 AP5E25 AP6A44 AP6E42 AP6L14 AP6M17 AP6T08
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAExtremity Medical - Disposable Orthopedic Surgical Instrument - 1.6 mm Guidewire - Reference Number 101-00006March 27, 2013
- FDA45® Angle Driver / Z-H, Greatbatch Medical 10 000 Wahrie Drive, Clarance, NY 14031. Model #: P2661, P2662,February 6, 2013
- FDATUNNELOC ROUND HEAD INTERFERENCE SCREW 8MMX25MM Biomet Sports Medicine manufactures a variety of internalFebruary 6, 2013
- FDAEsteem Programmer Model 6001, US and ISA, Model 3001 Dell E54XX, part of the Esteem Totally ImplantableFebruary 6, 2013
- FDAIO FiX X-Post Reamer, Extremity Medical, www.extremitymedical.com; Customer Service: 888.499.0079 The X-PostDecember 5, 2012
- FDAIO Fix IntraOsseous Fixation System; Extremity Medical. www.extremitymedical.com; Customer Service:November 28, 2012
Questions about this recall
Was Extremity Medical, IO FiX System Accessory. Catalog Number 101-00010. This screw measuring device is an accessory to the IO FiX Screw and Washer System that is used to measure the length of the screw by placing the depth gauge over the guidewire and down to the bone. Orthopedic device. recalled?
Yes. FDA published recall Z-0752-2013 on February 6, 2013. Reported hazard: Lag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical Technique Rev C. which can lead to possible inappropriate implant size selection (e.g. shorter screw size selected).
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail