FDAFDA · Food and drugs·Recall number Z-0753-2015

PKG, 5MM X 33CM INSERT, JOHANN FORCEPS, P/N 0Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. 250080763

Published December 24, 2014 · Recall initiated: November 17, 2014

Summary

FDA records medical device Class II recall Z-0753-2015, published December 24, 2014 and initiated November 17, 2014. Recalling firm: Stryker Endoscopy; Status: Terminated; Reported quantity: 180,573 units total; Lot: 641377 740261 840350 842445 843941 844410 847470 847622 940151 1040522 1040929 1041620 1042781 1042782 1042783 1042784 1240139 1241417 1241935 1242946 1245523 1245815 1246140 1246157 1340471 1341514 1342142 1342745 1343419 1345037 1345095 1345517 1346689 1441559 1442533 1443404 641681. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Endoscopy
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lot
641377 740261 840350 842445 843941 844410 847470 847622 940151 1040522 1040929 1041620 1042781 1042782 1042783 1042784 1240139 1241417 1241935 1242946 1245523 1245815 1246140 1246157 1340471 1341514 1342142 1342745 1343419 1345037 1345095 1345517 1346689 1441559 1442533 1443404 641681
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was PKG, 5MM X 33CM INSERT, JOHANN FORCEPS, P/N 0Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. 250080763 recalled?

Yes. FDA published recall Z-0753-2015 on December 24, 2014. Reported hazard: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

What should I do if I have this product?

Voluntary: Firm initiated · Letter