Cardinal Health Suction Canister Liner, MEDI-VAC CRD 1000 mL Suction Canister Liner Kit with 6 mm 1.8 m Tube, REF 65651-517
Published January 24, 2024 · Recall initiated: December 21, 2023
Summary
FDA records medical device Class II recall Z-0756-2024, published January 24, 2024 and initiated December 21, 2023. Recalling firm: Cardinal Health 200, LLC; Status: Ongoing; Reported quantity: 74515; Lot: 50630140152670 (case), Lot Numbers: J308-625, J308-629, J308-637, J309-616, J309-634, J309-646, J309-647, J310-603, J310-618. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cardinal Health 200, LLC
- Distributor
- US, Canada, Australia and New Zealand
- Lot
- 50630140152670 (case), Lot Numbers: J308-625, J308-629, J308-637, J309-616, J309-634, J309-646, J309-647, J310-603, J310-618
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health Presource PBDS kits, Containing BD Alaris Pump Infusion Sets used with BD Alaris Pump ModuleNovember 11, 2026
- FDACardinal Health MEDI-VAC CRD Suction Canister Liner, 1000 mL Suction Canister Liner Kit with 5 mm 1.8 m Tube,January 24, 2024
- FDAMaquet PowerLEDII OR Light System, Model Names and Catalog Numbers PWDII77SF - ARDPWT629100A, PWDII555DF -January 24, 2024
Questions about this recall
Was Cardinal Health Suction Canister Liner, MEDI-VAC CRD 1000 mL Suction Canister Liner Kit with 6 mm 1.8 m Tube, REF 65651-517 recalled?
Yes. FDA published recall Z-0756-2024 on January 24, 2024. Reported hazard: The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.
What should I do if I have this product?
Voluntary: Firm initiated · Letter