Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
Published February 6, 2019 · Recall initiated: December 21, 2018
Summary
FDA records medical device Class I recall Z-0763-2019, published February 6, 2019 and initiated December 21, 2018. Recalling firm: Edwards Lifesciences, LLC; Status: Terminated; Reported quantity: 875 units; Lot: Lot Numbers: 61321241, 61176369, 61176374, 61311580, 61176373, 61321254. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Edwards Lifesciences, LLC
- Distributor
- Worldwide Distribution
- Lot
- Lot Numbers: 61321241, 61176369, 61176374, 61311580, 61176373, 61321254
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGeneric Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery SystemNovember 11, 2026
- FDAEdwards EVOQUE tricuspid delivery system REF:9850TDS Sterile EOMay 20, 2026
- FDABrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS SoftwareJuly 16, 2025
- FDABrand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16July 16, 2025
- FDAPlums sold at ALDI, packaged in a 2-pound bag with the brand Rio Duero, EAN# 7804650090304.February 6, 2019
- FDAEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz ThermodilutionFebruary 6, 2019
Questions about this recall
Was Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction. recalled?
Yes. FDA published recall Z-0763-2019 on February 6, 2019. Reported hazard: These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.
What should I do if I have this product?
Voluntary: Firm initiated · Letter