Back to search
FDAFDA · Food and drugs·Recall number Z-0766-2013

P10 Tobii battery box (product number 510410) Product Usage - Battery pack for MyTobii P10 Device. MyTobii P10 is a portable eye-controlled communication device featuring the Tobii Eye Control system.

Published February 13, 2013 · Recall initiated: January 14, 2013

Summary

FDA records medical device Class II recall Z-0766-2013, published February 13, 2013 and initiated January 14, 2013. Recalling firm: Tobii Assistive Technology, Inc.; Status: Terminated; Reported quantity: 483 units; Lot: Product Code 510410. All units. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Tobii ATI customers P10 external battery pack, which contains sealed lead-acid batteries, malfunctioned while recharging and produced smoke with a chemical smell.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Tobii Assistive Technology, Inc.
Distributor
US Nationwide Distribution
Lot
Product Code 510410. All units.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was P10 Tobii battery box (product number 510410) Product Usage - Battery pack for MyTobii P10 Device. MyTobii P10 is a portable eye-controlled communication device featuring the Tobii Eye Control system. recalled?

Yes. FDA published recall Z-0766-2013 on February 13, 2013. Reported hazard: Tobii ATI customers P10 external battery pack, which contains sealed lead-acid batteries, malfunctioned while recharging and produced smoke with a chemical smell.

What should I do if I have this product?

Voluntary: Firm initiated · Letter