MCA7-1SB - 1MM Fluted Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.
Published January 22, 2014 · Recall initiated: March 29, 2012
Summary
FDA records medical device Class II recall Z-0766-2014, published January 22, 2014 and initiated March 29, 2012. Recalling firm: The Anspach Effort, Inc.; Status: Terminated; Reported quantity: 6439 total; Lot: All lots are included in recall. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- The Anspach Effort, Inc.
- Distributor
- Worldwide Distribution: US (nationwide) and Internationally to: AU, BE, CA, CH, ES, FR, GB, IT, JP, NL, SE, and ZA.
- Lot
- All lots are included in recall
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAElectric Handpiece EM Hand Control Product Usage: Electric systemMay 10, 2017
- FDA14cm QD Angle Attachment Product Usage: Pneumatic systemMay 10, 2017
- FDAKeyless Driver Product Usage: Pneumatic systemMay 10, 2017
- FDA6.8cm Controlled Depth Attachment Product Usage: Pneumatic systemMay 10, 2017
- FDAMCA7-15SD - 1.5MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting andJanuary 22, 2014
- FDAMCA7-1SD - 1MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shapingJanuary 22, 2014
Questions about this recall
Was MCA7-1SB - 1MM Fluted Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy. recalled?
Yes. FDA published recall Z-0766-2014 on January 22, 2014. Reported hazard: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
What should I do if I have this product?
Voluntary: Firm initiated · Letter