FDAFDA · Food and drugs·Recall number Z-0767-2025

Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;

Published January 1, 2025 · Recall initiated: November 12, 2024

Summary

FDA records medical device Class II recall Z-0767-2025, published January 1, 2025 and initiated November 12, 2024. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Ongoing; Reported quantity: 13,115 units (11,520 US, 1,595 ROW); Lot: Siemens Material Number (SMN): 10445158; Catalog Number: K3057; UDI-DI: 00842768015694; Lot numbers: 24057BB, 24064BC. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista® Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Slovenia, Spain.
Lot
Siemens Material Number (SMN): 10445158; Catalog Number: K3057; UDI-DI: 00842768015694; Lot numbers: 24057BB, 24064BC.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158; recalled?

Yes. FDA published recall Z-0767-2025 on January 1, 2025. Reported hazard: Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista® Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out

What should I do if I have this product?

Voluntary: Firm initiated · Letter