Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
Published December 21, 2016 · Recall initiated: September 16, 2016
Summary
FDA records medical device Class II recall Z-0769-2017, published December 21, 2016 and initiated September 16, 2016. Recalling firm: Natus Medical Incorporated; Status: Terminated; Reported quantity: 713 units total (403 units in US and 310 units outside US); Lot: *************Serial Numbers of Units distributed in the US ************** 004329, 004330, 004349, 004360, 004361, 004362, 004363, 004364, 004365, 004366, 004367, 004369, 004370, 004371, 004417, 004421, 004426, 004436, 004437, 004443, 004448, 004449, 004454, 004456, 004473, 004481, 004488, 004490, 004494, 004495, 004502, 004507, 004517, 004523, 004525, 004526, 004527, 004528, 004530, 004533, 004539, 004541, 004544, 004545, 004546, 004548, 004552, 004553, 004554, 004555, 004556, 004557, 004558, 004567, 004568, 004573, 004578, 004581, 004582, 004584, 004588, 004591, 004602, 004608, 004630, 004632, 004637, 004640, 004647, 004659, 004667, 004668, 004670, 004672, 004673, 004677, 004678, 004679, 004681, 004683, 004684, 004685, 004686, 004693, 004694, 004696, 004700, 004706, 004707, 004708, 004709, 004710, 004711, 004712, 004713, 004714, 004715, 004716, 004717, 004718, 004719, 004720, 004721, 004722, 004723, 004724, 004725, 004726, 004727, 004728, 004729, 004730, 004731, 004732, 004733, 004734, 004735, 004736, 004737, 004738, 004739, 004740, 004741, 004742, 004743, 004745, 004746, 004747, 004749, 004750, 004753, 004754, 004756, 004757, 004760, 004761, 004762, 004763, 004764, 004766, 004767, 004768, 004771, 004772, 004773, 004774, 004775, 004776, 004777, 004778, 004779, 004780, 004781, 004782, 004783, 004784, 004785, 004787, 004788, 004789, 004790, 004791, 004793, 004794, 004795, 004796, 004797, 004798, 004800, 004801, 004802, 004803, 004804, 004805, 004806, 004807, 004808, 004810, 004811, 004812, 004814, 004815, 004816, 004818, 004819, 004820, 004822, 004823, 004824, 004825, 004826, 004827, 004828, 004829, 004830, 004831, 004832, 004833, 004834, 004835, 004836, 004837, 004838, 004839, 004840, 004841, 004842, 004843, 004844, 004845, 004846, 004847, 004848, 004849, 004850, 004851, 004852, 004853, 004854, 004855, 004856, 004857, 004858, 004859, 004860, 004861, 004862, 004863, 004864, 004865, 004866, 004867, 004868, 004869, 004870, 004871, 004872, 004873, 004874, 004876, 004877, 004878, 004879, 004880, 004881, 004882, 004884, 004885, 004886, 004887, 004888, 004890, 004891, 004893, 004894, 004895, 004896, 004897, 004899, 004900, 004901, 004902, 004903, 004904, 004905, 004906, 004907, 004909, 004910, 004997, 004998, 004999, 005007, 005008, 005009, 005010, 005011, 005012, 005013, 005014, 005015, 005016, 005017, 005018, 005019, 005020, 005021, 005022, 005023, 005024, 005025, 005026, 005027, 005028, 005029, 005030, 005031, 005032, 005033, 005034, 005035, 005036, 005037, 005039, 005040, 005041, 005042, 005043, 005044, 005045, 005046, 005047, 005048, 005049, 005050, 005051, 005052, 005053, 005054, 005055, 005056, 005057, 005058, 005059, 005060, 005061, 005062, 005063, 005064, 005065, 005066, 005067, 005068, 005069, 005070, 005071, 005072, 005073, 005074, 005075, 005076, 005077, 005078, 005079, 005080, 005081, 005082, 005083, 005084, 005085, 005086, 005087, 005088, 005089, 005090, 005091, 005092, 005093, 005094, 005095, 005096, 005097, 005098, 005099, 005100, 005101, 005103, 005104, 005105, 005106, 005107, 005108, 005112, 005113, 005115, 005117, 005118, 005119, 005120, 005121, 005122, 005123, 005124, 005125, 005128, 005136, 005137, 005138, 005139, 005140, 005141, 005142, 005143, 005144, 005145, 005146, 005147, 005148, 005149, 005150, 005151, 005152, and 005153. *************Serial Numbers of Units distributed in the US ************** 004283, 004284, 004285, 004286, 004287, 004288, 004290, 004291, 004292, 004293, 004294, 004295, 004296, 004298, 004299, 004302, 004303, 004305, 004306, 004307, 004308, 004309, 004310, 004312, 004313, 004314, 004315, 004316, 004319, 004320, 004321, 004322, 004325, 004326, 004327, 004328, 004332, 004333, 004334, 004335, 004337, 004338, 004339, 004340, 004341, 004343, 004344, 004345, 004347, 004348, 004350, 004351, 004353, 004354, 004355, 004356, 004357, 004359, 004372, 004373, 004374, 004375, 004379, 004380, 004381, 004382, 004383, 004384, 004385, 004386, 004388, 004389, 004390, 004391, 004392, 004393, 00. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Natus Medical Incorporated
- Distributor
- Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, NC, ND, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, WV and Puerto Rico. , and to the countries of : Argentina, Austria, Canada, Chile, China, Denmark, Dominican Republic, Finland, France, Germany, Hong Kong, Hungary, Indonesia, Italy, Latvia, Malaysia, Namibia, Paraguay, Poland, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Taiwan, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Virgin Islands (British).
- Lot
- *************Serial Numbers of Units distributed in the US ************** 004329, 004330, 004349, 004360, 004361, 004362, 004363, 004364, 004365, 004366, 004367, 004369, 004370, 004371, 004417, 004421, 004426, 004436, 004437, 004443, 004448, 004449, 004454, 004456, 004473, 004481, 004488, 004490, 004494, 004495, 004502, 004507, 004517, 004523, 004525, 004526, 004527, 004528, 004530, 004533, 004539, 004541, 004544, 004545, 004546, 004548, 004552, 004553, 004554, 004555, 004556, 004557, 004558, 004567, 004568, 004573, 004578, 004581, 004582, 004584, 004588, 004591, 004602, 004608, 004630, 004632, 004637, 004640, 004647, 004659, 004667, 004668, 004670, 004672, 004673, 004677, 004678, 004679, 004681, 004683, 004684, 004685, 004686, 004693, 004694, 004696, 004700, 004706, 004707, 004708, 004709, 004710, 004711, 004712, 004713, 004714, 004715, 004716, 004717, 004718, 004719, 004720, 004721, 004722, 004723, 004724, 004725, 004726, 004727, 004728, 004729, 004730, 004731, 004732, 004733, 004734, 004735, 004736, 004737, 004738, 004739, 004740, 004741, 004742, 004743, 004745, 004746, 004747, 004749, 004750, 004753, 004754, 004756, 004757, 004760, 004761, 004762, 004763, 004764, 004766, 004767, 004768, 004771, 004772, 004773, 004774, 004775, 004776, 004777, 004778, 004779, 004780, 004781, 004782, 004783, 004784, 004785, 004787, 004788, 004789, 004790, 004791, 004793, 004794, 004795, 004796, 004797, 004798, 004800, 004801, 004802, 004803, 004804, 004805, 004806, 004807, 004808, 004810, 004811, 004812, 004814, 004815, 004816, 004818, 004819, 004820, 004822, 004823, 004824, 004825, 004826, 004827, 004828, 004829, 004830, 004831, 004832, 004833, 004834, 004835, 004836, 004837, 004838, 004839, 004840, 004841, 004842, 004843, 004844, 004845, 004846, 004847, 004848, 004849, 004850, 004851, 004852, 004853, 004854, 004855, 004856, 004857, 004858, 004859, 004860, 004861, 004862, 004863, 004864, 004865, 004866, 004867, 004868, 004869, 004870, 004871, 004872, 004873, 004874, 004876, 004877, 004878, 004879, 004880, 004881, 004882, 004884, 004885, 004886, 004887, 004888, 004890, 004891, 004893, 004894, 004895, 004896, 004897, 004899, 004900, 004901, 004902, 004903, 004904, 004905, 004906, 004907, 004909, 004910, 004997, 004998, 004999, 005007, 005008, 005009, 005010, 005011, 005012, 005013, 005014, 005015, 005016, 005017, 005018, 005019, 005020, 005021, 005022, 005023, 005024, 005025, 005026, 005027, 005028, 005029, 005030, 005031, 005032, 005033, 005034, 005035, 005036, 005037, 005039, 005040, 005041, 005042, 005043, 005044, 005045, 005046, 005047, 005048, 005049, 005050, 005051, 005052, 005053, 005054, 005055, 005056, 005057, 005058, 005059, 005060, 005061, 005062, 005063, 005064, 005065, 005066, 005067, 005068, 005069, 005070, 005071, 005072, 005073, 005074, 005075, 005076, 005077, 005078, 005079, 005080, 005081, 005082, 005083, 005084, 005085, 005086, 005087, 005088, 005089, 005090, 005091, 005092, 005093, 005094, 005095, 005096, 005097, 005098, 005099, 005100, 005101, 005103, 005104, 005105, 005106, 005107, 005108, 005112, 005113, 005115, 005117, 005118, 005119, 005120, 005121, 005122, 005123, 005124, 005125, 005128, 005136, 005137, 005138, 005139, 005140, 005141, 005142, 005143, 005144, 005145, 005146, 005147, 005148, 005149, 005150, 005151, 005152, and 005153. *************Serial Numbers of Units distributed in the US ************** 004283, 004284, 004285, 004286, 004287, 004288, 004290, 004291, 004292, 004293, 004294, 004295, 004296, 004298, 004299, 004302, 004303, 004305, 004306, 004307, 004308, 004309, 004310, 004312, 004313, 004314, 004315, 004316, 004319, 004320, 004321, 004322, 004325, 004326, 004327, 004328, 004332, 004333, 004334, 004335, 004337, 004338, 004339, 004340, 004341, 004343, 004344, 004345, 004347, 004348, 004350, 004351, 004353, 004354, 004355, 004356, 004357, 004359, 004372, 004373, 004374, 004375, 004379, 004380, 004381, 004382, 004383, 004384, 004385, 004386, 004388, 004389, 004390, 004391, 004392, 004393, 00
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAExternal Drainage System Collection Bag, REF: NT821732C, STERILEEO, Rx onlyJanuary 10, 2024
- FDAEDS 3 CSF External Drainage System, REF: NT821731C, STERILEEO, Rx onlyJanuary 10, 2024
- FDAThe Olympic Pasteurmatic 3000 and Olympic Pasteurmatic 3500 are intended for cleaning and high-levelJuly 19, 2017
- FDASecond generation neoBLUE LED Phototherapy Light (neoBLUE 2) sold between 2003 and 2008. Catalog Part NumberMay 10, 2017
- FDAMicroScan LabPro Data Management System. Intended to manage both microbial identification (ID) andDecember 21, 2016
- FDANatus neoBLUE blanket LED Phototherapy system, neonatal phototherapy. Used to provide treatment for neonatalDecember 21, 2016
Questions about this recall
Was Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS recalled?
Yes. FDA published recall Z-0769-2017 on December 21, 2016. Reported hazard: The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.
What should I do if I have this product?
Voluntary: Firm initiated · Letter