Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image and interpretations. Mammongraphic may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.
Published January 22, 2014 · Recall initiated: March 4, 2011
Summary
FDA records medical device Class II recall Z-0773-2014, published January 22, 2014 and initiated March 4, 2011. Recalling firm: Ge Healthcare It; Status: Terminated; Reported quantity: 43 units; Lot: Imagecast PACS versions prior to 3.6.51.01. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing. Because the estimated magnification factor is not used, distance meas
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ge Healthcare It
- Distributor
- Nationwide Distribution: US including states of: AL, AZ, CT, FL, GA, IA, IN, LA, MA, MI, MN, MO, NC, NH, NJ, NY, OH, PA, TX, VA, WI, WV + District of Columbia.
- Lot
- Imagecast PACS versions prior to 3.6.51.01
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity Enterprise Archive is a software product for receiving, archiving and sending of medical data.December 31, 2014
- FDACentricity PACS System versions 3.X and higher and 4.X and higher Product Usage: The Centricity PACSJanuary 22, 2014
- FDABD Interlink threaded lock cannula This is a locking blunt plastic cannula specifically designed for IVJanuary 22, 2014
- FDAEstrogen Receptor AB-11 (Clone ID5) Mouse Monoclonal Antibody 1 mk (2mg.ml) Manufactured by: Lab VisionJanuary 22, 2014
- FDAThe GE Centricity Web; Software Version 3.0x, is intended for use under the direct supervision of a licensedDecember 11, 2013
- FDACentricity PACS IW by GE Healthcare Dynamic Imaging Solutions is a device that receives medical imagesNovember 6, 2013
Questions about this recall
Was Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image and interpretations. Mammongraphic may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA. recalled?
Yes. FDA published recall Z-0773-2014 on January 22, 2014. Reported hazard: In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing. Because the estimated magnification factor is not used, distance meas
What should I do if I have this product?
Voluntary: Firm initiated · Letter