Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.

Published February 13, 2013 · Recall initiated: September 5, 2012

Summary

FDA records medical device Class II recall Z-0783-2013, published February 13, 2013 and initiated September 5, 2012. Recalling firm: The Anspach Effort, Inc.; Status: Terminated; Reported quantity: 8 devices; Lot: Lot number: 2000930. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
The Anspach Effort, Inc.
Distributor
Distributed in the state of Massachusetts
Lot
Lot number: 2000930
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures. recalled?

Yes. FDA published recall Z-0783-2013 on February 13, 2013. Reported hazard: Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.

What should I do if I have this product?

Voluntary: Firm initiated · Letter