Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures.
Published February 13, 2013 · Recall initiated: September 5, 2012
Summary
FDA records medical device Class II recall Z-0783-2013, published February 13, 2013 and initiated September 5, 2012. Recalling firm: The Anspach Effort, Inc.; Status: Terminated; Reported quantity: 8 devices; Lot: Lot number: 2000930. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- The Anspach Effort, Inc.
- Distributor
- Distributed in the state of Massachusetts
- Lot
- Lot number: 2000930
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAElectric Handpiece EM Hand Control Product Usage: Electric systemMay 10, 2017
- FDA14cm QD Angle Attachment Product Usage: Pneumatic systemMay 10, 2017
- FDAKeyless Driver Product Usage: Pneumatic systemMay 10, 2017
- FDA6.8cm Controlled Depth Attachment Product Usage: Pneumatic systemMay 10, 2017
- FDASynthes(R) Spine Vectra, Vectra-T and Vectra-One Anterior Cervical Plating for Spinal Fusion, Adjustable DrillFebruary 13, 2013
- FDATHERA PEARL 3-in-1 BREAST THERAPY, 2 hot and cold packs in plastic box. The Thera Pearl 3-in-1 Breast TherapyFebruary 13, 2013
Questions about this recall
Was Anspach Effort, Inc., 98-0061 MA-15C Bearing Sleeve with Removable Bur Guard. A reusable device used with dissection tools; designed for Transphenoidal and Skull base procedures. recalled?
Yes. FDA published recall Z-0783-2013 on February 13, 2013. Reported hazard: Contact between the bur and bur guard could generate metal fragments that may or may not be visible to the surgeon and can potentially remain in the surgical site. The materials used to fabricate the bur guard are not traceable, design validation did not effectively evaluate adequate protection of adjacent tissue and inspection results for each of these bur guards were not documented.
What should I do if I have this product?
Voluntary: Firm initiated · Letter