Zimmer Segmental Systems; Polyethylene Inserts, Size B. Designed to address significant bone loss resulting from oncology, trauma, and/or salvage of previously failed arthroplasty.
Published January 29, 2014 · Recall initiated: December 19, 2013
Summary
FDA records medical device Class II recall Z-0783-2014, published January 29, 2014 and initiated December 19, 2013. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 5305 distribution events; Lot: Part numbers 00-5850-012-95. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution. USA nationwide and Germany, with further distribution into EU countries.
- Lot
- Part numbers 00-5850-012-95
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDAUP & UP kids' power toothbrushJanuary 29, 2014
- FDAZimmer Segmental Systems; Polyethylene Inserts, Size C. Knee replacement.January 29, 2014
Questions about this recall
Was Zimmer Segmental Systems; Polyethylene Inserts, Size B. Designed to address significant bone loss resulting from oncology, trauma, and/or salvage of previously failed arthroplasty. recalled?
Yes. FDA published recall Z-0783-2014 on January 29, 2014. Reported hazard: After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail