Back to search
FDAFDA · Food and drugs·Recall number Z-0789-2022

SAM CHEST SEAL REF CS203-EN Combo LOT Y060321-09

Published March 23, 2022 · Recall initiated: January 28, 2022

Summary

FDA records medical device Class II recall Z-0789-2022, published March 23, 2022 and initiated January 28, 2022. Recalling firm: Sam Medical Products; Status: Terminated; Reported quantity: 2,950 packages; Lot: Model Number: CS203-EN UDI Code: (01)10822045000190(10)Y060321-09(17)270601 (Case Carton) UDI Code: (01)10822045000193(10)Y060321-09(17)270601 (Each Pouch/Package) Lot Number: Y060321-09. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to partially assembled valved dressing that has a hole within it instead of the intended non-valved dressing in the package resulting in that the dressing cannot be centered over the open chest wound and create an occlusive seal due to the hole being present.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Sam Medical Products
Distributor
U.S.: AZ, CA, CO, FL, GA, MA, NY, OR, PA, Puerto Rico, RI, TX, VA, and WA O.U.S.: Canada
Lot
Model Number: CS203-EN UDI Code: (01)10822045000190(10)Y060321-09(17)270601 (Case Carton) UDI Code: (01)10822045000193(10)Y060321-09(17)270601 (Each Pouch/Package) Lot Number: Y060321-09
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SAM CHEST SEAL REF CS203-EN Combo LOT Y060321-09 recalled?

Yes. FDA published recall Z-0789-2022 on March 23, 2022. Reported hazard: Due to partially assembled valved dressing that has a hole within it instead of the intended non-valved dressing in the package resulting in that the dressing cannot be centered over the open chest wound and create an occlusive seal due to the hole being present.

What should I do if I have this product?

Voluntary: Firm initiated · Letter