Fresenius Liberty Cycler Set Single Conn. Ext. DL. Part Number: 050-87216. Acute and chronic peritoneal dialysis.
Published February 20, 2013 · Recall initiated: October 12, 2012
Summary
FDA records medical device Class II recall Z-0799-2013, published February 20, 2013 and initiated October 12, 2012. Recalling firm: Fresenius Medical Care Holdings, Inc.; Status: Terminated; Reported quantity: 560 cases, totaling 5590 units; Lot: Lot 12BR08009. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a small risk of a leak in Cycler Sets and if the Cycler Set is punctured, the dialysate may become contaminated potentially resulting in peritonitis.
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Fresenius Medical Care Holdings, Inc.
- Distributor
- Nationwide Distribution.
- Lot
- Lot 12BR08009
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA5008X Blood Tubing Set, 24 units per case. Includes the below model numbers and descriptions. 1. 5008X HDJuly 15, 2026
- FDABicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;June 17, 2026
- FDABicarby Dialysate; Model number: RFP-400-G;June 17, 2026
- FDADELFLEX, Dextrose Peritoneal Dialysis Solution with attached stay-safe Exchange Set for IntraperitonealMay 13, 2026
- FDAMindray DS USA, Inc., DPM Central Monitoring System with ambulatory Telepack. Part numbers: 300BF-PA00002, andFebruary 20, 2013
- FDABacterin International, Inc., Osteoselect Demineralized Bone Matrix DMB Putty; Graft ID# B090056-664 Donor#February 20, 2013
Questions about this recall
Was Fresenius Liberty Cycler Set Single Conn. Ext. DL. Part Number: 050-87216. Acute and chronic peritoneal dialysis. recalled?
Yes. FDA published recall Z-0799-2013 on February 20, 2013. Reported hazard: There is a small risk of a leak in Cycler Sets and if the Cycler Set is punctured, the dialysate may become contaminated potentially resulting in peritonitis.
What should I do if I have this product?
Voluntary: Firm initiated · Telephone