DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray System is a mobile imaging system that incorporates a self contained x-ray generator, image receptor, imaging display and software for acquiring medical diagnostic images outside of a standard x-ray room. The DRX-Revolution Mobile X-Ray System has a motor for driving the system so that the operator does not have to push the system, but can walk behind the system while it is moving from one location to another.

Published December 24, 2014 · Recall initiated: November 25, 2014

Summary

FDA records medical device Class II recall Z-0800-2015, published December 24, 2014 and initiated November 25, 2014. Recalling firm: Carestream Health Inc; Status: Terminated; Reported quantity: 629 units (Domestic: 358 units; Foreign: 271 units); Lot: Serial Numbers 101 to 764. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Carestream Health Inc
Distributor
Worldwide Distribution: US (nationwide) and countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, ITALY, JAPAN, KOREA, KUWAIT, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, and UNITED KINGDOM.
Lot
Serial Numbers 101 to 764
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray System is a mobile imaging system that incorporates a self contained x-ray generator, image receptor, imaging display and software for acquiring medical diagnostic images outside of a standard x-ray room. The DRX-Revolution Mobile X-Ray System has a motor for driving the system so that the operator does not have to push the system, but can walk behind the system while it is moving from one location to another. recalled?

Yes. FDA published recall Z-0800-2015 on December 24, 2014. Reported hazard: While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray

What should I do if I have this product?

Voluntary: Firm initiated · Letter