Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.
Published February 20, 2013 · Recall initiated: December 17, 2012
Summary
FDA records medical device Class II recall Z-0810-2013, published February 20, 2013 and initiated December 17, 2012. Recalling firm: Mizuho OSI; Status: Terminated; Reported quantity: 869 devices; Lot: Table model 5855, sub assembly 5855-901 Orthopedic Table Spar, manufactured between 1996 and 2007. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Mizuho OSI
- Distributor
- USA Distribution -- Worldwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.
- Lot
- Table model 5855, sub assembly 5855-901 Orthopedic Table Spar, manufactured between 1996 and 2007.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA216 Wilson Frame Pads, Model # 5319-37; Used with Radiolucent Wilson Frames, model numbers 5319G and 5321G.April 26, 2017
- FDAChenica Gunnell Rehab Series of Wheelchairs (GRSW); Mechanical Wheelchair, Sold as Rehab Tough and Tilt (TNT),February 20, 2013
- FDAUltradent Products, Chromaclone PVS, Regular Set heavy body, Bubble Gum Flavor, 50 ml, Impression Material,February 20, 2013
- FDAOxoid Antimicrobial Susceptibility Discs, Cefoxitin 30ug REF CT0119B, packaged with 50 discs per cartridge, 5February 20, 2013
- FDAUltradent Products, Chromaclone PVS, Fast Set heavy body, Bubble Gum Flavor, 50 ml, Impression Material, PartFebruary 20, 2013
- FDARemel Xpect Clostridium difficile Toxin A/B, IVD, 40 clostridium difficile toxin A/B tests, REF 24640. ProductFebruary 20, 2013
Questions about this recall
Was Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar. recalled?
Yes. FDA published recall Z-0810-2013 on February 20, 2013. Reported hazard: Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.
What should I do if I have this product?
Voluntary: Firm initiated · Letter