Microbiologics QC Sets and Panels, Gram-Positive Blood Culture Control Panel (Inactivated Pellet), Catalog # 8180 Used in test systems to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation.
Published January 22, 2020 · Recall initiated: December 16, 2019
Summary
FDA records medical device Class II recall Z-0812-2020, published January 22, 2020 and initiated December 16, 2019. Recalling firm: Microbiologics Inc; Status: Terminated; Reported quantity: 810; Lot: Lot # 8180-46 8180-47 8180-48 8180-49 8180-50 8180-51 8180-52. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The recommended hydration fluid for the 8180 Gram-Positive Blood Culture Control Panel (Inactivated Pellet), has changed from sterile 0.85% saline or sterile blood culture media to sterile DI water or pH 7.2 phosphate buffer.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Microbiologics Inc
- Distributor
- Distributed nationwide including Washington DC, and distributed internationally to Germany and Netherlands.
- Lot
- Lot # 8180-46 8180-47 8180-48 8180-49 8180-50 8180-51 8180-52
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAKWIK-STIK, Quality control kit for culture media, Vibrio parahaemolyticus derived from ATCC 17802, CatalogDecember 10, 2025
- FDALYFO-DISK, M. Canis Derived from ATCC 36299, packaged as: 1) REF 0894K (6-pack), 2) REF 0894P (2-pack)August 6, 2025
- FDAKWIK-STICK, Quality control kit for culture media, Catalog Numbers 01065K, Oligella urethralis derived fromJune 18, 2025
- FDAMicrobiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235;February 19, 2025
- FDASiemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System - Product Usage: AngiographyJanuary 22, 2020
- FDADawson-Mueller Drainage Catheter, Multipurpose Drainage Catheter, RPN UL T6.3-35-15-P-5S-CLDM-HC, UL T6.January 22, 2020
Questions about this recall
Was Microbiologics QC Sets and Panels, Gram-Positive Blood Culture Control Panel (Inactivated Pellet), Catalog # 8180 Used in test systems to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. recalled?
Yes. FDA published recall Z-0812-2020 on January 22, 2020. Reported hazard: The recommended hydration fluid for the 8180 Gram-Positive Blood Culture Control Panel (Inactivated Pellet), has changed from sterile 0.85% saline or sterile blood culture media to sterile DI water or pH 7.2 phosphate buffer.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail