Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: AHN BIOPSY STERILE, kit number AHBP12A

Published February 7, 2024 · Recall initiated: November 29, 2023

Summary

FDA records medical device Class II recall Z-0821-2024, published February 7, 2024 and initiated November 29, 2023. Recalling firm: American Contract Systems, Inc.; Status: Ongoing; Reported quantity: 95 units; Lot: AHBP12A, UDI/DI 00191072188229, Lot Numbers: 963231, exp. 1/21/2024; 940231, exp. 1/21/2024; 996231, exp. 2/28/2023. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
American Contract Systems, Inc.
Distributor
US Nationwide distribution.
Lot
AHBP12A, UDI/DI 00191072188229, Lot Numbers: 963231, exp. 1/21/2024; 940231, exp. 1/21/2024; 996231, exp. 2/28/2023
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: AHN BIOPSY STERILE, kit number AHBP12A recalled?

Yes. FDA published recall Z-0821-2024 on February 7, 2024. Reported hazard: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail